Getinge 88-Series Washer- Disinfector. Recall
Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Getinge 88-Series Washer- Disinfector.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0016-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 29, 2025
- FDA report date
- Oct 8, 2025
- Quantity in commerce
- 902 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
Affected model & UDI checker
Search the exact identifiers published for Z-0016-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 88-5 | Getinge 88-Series Washer- Disinfector. | Check source notice | Not specified |
| Not specified | Getinge 88-Series Washer- Disinfector. | Check source notice | 07340153700109 |
| Not specified | Getinge 88-Series Washer- Disinfector. | Check source notice | 07340153700116 |
FDA code information
Model Number: 88-5. UDI-DI Numbers: 07340153700109, 07340153700116. Software Version: GD 14080 Rev A. All Lot Numbers with electrical heating.
Why was this product recalled?
Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Getinge Disinfection Ab take?
Getinge notified consignees on about 08/29/2025 via letter. Consignees were instructed that Getinge service technicians will be proactively reaching out to ensure affected devices are calibrated appropriately. In the meantime, Getinge has developed a verification procedure to ensure that the circulation pressure in standby mode is correctly calibrated to 0.0 kPa. If the value deviates, recalibration by a qualified technician is required. The verification of the circulation pressure shall be performed by the operator or Getinge technician. If the value deviates from 0.0 kPa out of allowed tolerances (1 kPa), stop using the device immediately and work with your Getinge service technician to arrange recalibration. Consignees were also instructed to ensure all caregivers and users of the affected product are made aware of the recall and to complete and return the provided response form. ***UPDATE 10/16/2025: Firm notified FDA of an expansion to include additional OUS units and one additional complaint. ***UPDATE 03/02/2026: Firm notified FDA of an expansion to include additional OUS units and one additional complaint.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0016-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology