FDA RECALL RECORD · Z-0016-2026

Getinge 88-Series Washer- Disinfector. Recall

Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Getinge 88-Series Washer- Disinfector.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0016-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 29, 2025
FDA report date
Oct 8, 2025
Quantity in commerce
902 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

Affected model & UDI checker

Search the exact identifiers published for Z-0016-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
88-5Getinge 88-Series Washer- Disinfector.Check source noticeNot specified
Not specifiedGetinge 88-Series Washer- Disinfector.Check source notice07340153700109
Not specifiedGetinge 88-Series Washer- Disinfector.Check source notice07340153700116
FDA code information

Model Number: 88-5. UDI-DI Numbers: 07340153700109, 07340153700116. Software Version: GD 14080 Rev A. All Lot Numbers with electrical heating.

View the original FDA recall notice

Why was this product recalled?

Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Getinge Disinfection Ab take?

Getinge notified consignees on about 08/29/2025 via letter. Consignees were instructed that Getinge service technicians will be proactively reaching out to ensure affected devices are calibrated appropriately. In the meantime, Getinge has developed a verification procedure to ensure that the circulation pressure in standby mode is correctly calibrated to 0.0 kPa. If the value deviates, recalibration by a qualified technician is required. The verification of the circulation pressure shall be performed by the operator or Getinge technician. If the value deviates from 0.0 kPa out of allowed tolerances (1 kPa), stop using the device immediately and work with your Getinge service technician to arrange recalibration. Consignees were also instructed to ensure all caregivers and users of the affected product are made aware of the recall and to complete and return the provided response form. ***UPDATE 10/16/2025: Firm notified FDA of an expansion to include additional OUS units and one additional complaint. ***UPDATE 03/02/2026: Firm notified FDA of an expansion to include additional OUS units and one additional complaint.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.