Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0022-2026

NES Reprocessed 0.9mm RX Turbo Elite Laser Atherectomy Catheter. Recall

NES Reprocessed 0.9mm RX Turbo Elite Laser Atherectomy Catheter. Model/Catalog Number: R-410-154. The NES Reprocessed Turbo-Elite Laser Atherectomy Catheter is indicated for use in the treatment, including atherectomy, of infrainguinal stenoses and occlusions. This model is not cleared for marketing in the US.

FDA recall Class II

Ongoing

FDA lists this Class II recall for NES Reprocessed 0.9mm RX Turbo Elite Laser Atherectomy Catheter.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0022-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 29, 2025
FDA report date
Oct 8, 2025
Quantity in commerce
795 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

NES Reprocessed 0.9mm RX Turbo Elite Laser Atherectomy Catheter. Model/Catalog Number: R-410-154. The NES Reprocessed Turbo-Elite Laser Atherectomy Catheter is indicated for use in the treatment, including atherectomy, of infrainguinal stenoses and occlusions. This model is not cleared for marketing in the US.

Affected model & UDI checker

Search the exact identifiers published for Z-0022-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
R-410-154NES Reprocessed 0.9mm RX Turbo Elite Laser Atherectomy Catheter.Check source noticeNot specified
Not specifiedNES Reprocessed 0.9mm RX Turbo Elite Laser Atherectomy Catheter.Check source notice00850044399062
FDA code information

Model No R-410-154; UDI-DI 00850044399062; All Lot Numbers that are within their labeled shelf life with expiration dates prior to 29AUG2026.

View the original FDA recall notice

Why was this product recalled?

Potential for breaches in the sterile barrier packaging, compromising sterility assurance.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Northeast Scientific Inc. take?

Northeast Scientific notified consignees via an "URGENT: Medical Device Recall Notification" letter dated 8/29/2025 and email. Consignees were instructed to immediately discontinue use of affected products, quarantine and return any unused affected units, and complete and return the enclosed Recall Response Form, even if you no longer have any affected product. If product were further distributed, consignees were instructed to notify those customers immediately and ensure they also discontinue use and return the devices. Northeast Scientific will provide replacement product (if available) or credit for returned product. In addition, Northeast Scientific has initiated an investigation and is implementing corrective actions to prevent recurrence. If you have any questions or concerns, please reach out to your Northeast Scientific Sales Representative or contact us at Sales@nescientific.com or 203-756-2111.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology