Brand Name: Expressaire Product Name: DISPOSABLE TOURNIQUET CUFF, 12" X 3.5" 2-LINE (CS 12) PLUM Recall
Brand Name: Expressaire Product Name: DISPOSABLE TOURNIQUET CUFF, 12" X 3.5" 2-LINE (CS 12) PLUM Model/Catalog Number: A70107003 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Expressaire Product Name: DISPOSABLE TOURNIQUET CUFF, 12" X 3.5" 2-LINE (CS 12) PLUM. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0024-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 24, 2026
- FDA report date
- Oct 7, 2026
- Quantity in commerce
- 1899
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Brand Name: Expressaire Product Name: DISPOSABLE TOURNIQUET CUFF, 12" X 3.5" 2-LINE (CS 12) PLUM Model/Catalog Number: A70107003 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No
Affected model & UDI checker
Search the exact identifiers published for Z-0024-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| A70107003 | Brand Name: Expressaire Product Name: DISPOSABLE TOURNIQUET CUFF, 12" X 3.5" 2-LINE (CS 12) PLUM | Check source notice | Not specified |
FDA code information
Lot Code: Model Number: A70107003 UDI-DI: (01) 00885914035474 All Lots produced from 01SEP2016 to 01MAR2023: R0601STY, Q0301STY, Q1115STY, P0620STY, P1015STY, O0815STY, N0330STY, M0130STY, M0430STY, M0730STY, M0930STY, O0120STY
Why was this product recalled?
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
What actions did ASP Global, LLC. take?
On 08/24/2026, an "URGENT MEDICAL DEVICE RECALL" Letter was emailed to customers informing them that, ASP Global identified a discrepancy between the labeled shelf life and available shelf-life validation data for certain Sterile Disposable Tourniquet Cuff products. While affected products were labeled with a 10-year shelf life, available validation documentation supports a shelf life of 5-years. Customer are instructed to: a) Review Appendix A to identify affected item codes and lot numbers. b) Discontinue use of the affected products. c) Quarantine any remaining impacted inventory to prevent further use. d) Complete and return the enclosed Recall Response Form to confirm receipt and actions to FieldActions@aspglobal.com. e) Dispose remaining inventory in accordance with the local facility s guidelines or notify ASP Global at FieldActions@aspglobal.com for return authorization and disposal assistance. f) If customers have distributed affected products to other facilities or customers, notify those recipients of this recall and provide them with a copy of this notification. For questions regarding this recall, contact: ASP Global Field Actions Team Email: FieldActions@aspglobal.com Phone: (855)-867-2190| Mon Fri, 9:00 AM 5:00 PM EST
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0024-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology