FDA RECALL RECORD · Z-0028-2026

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I Recall

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

FDA recall Class I

Ongoing

FDA lists this Class I recall for NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0028-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Sep 9, 2025
FDA report date
Oct 15, 2025
Quantity in commerce
1667 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

Affected model & UDI checker

Search the exact identifiers published for Z-0028-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedNOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-ICheck source notice05060541640009
FDA code information

Model Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.

View the original FDA recall notice

Why was this product recalled?

Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).

What actions did NOXBOX LTD take?

On September 9, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: - Use the instructions provided to troubleshoot the device. - Complete and return the attached response form to Linde Gas & Equipment as instructed on the form. Recalling firm actions: NOxBOX Ltd. is in the process of developing a new software version that will include a modification to the software to resolve this issue, which will be released pending required regulatory review and approval. Once the software version is released, NOxBOX Ltd. will coordinate with Linde Gas & Equipment in the U.S. to update your device. If you have any questions about this notice or need assistance, please reach out to your local Clinical Educator or the Customer Excellence Center at 1-833-NOX-VENT (669-8368) or at LG.US.NOXIVENT@Linde.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.