FDA RECALL RECORD · Z-0029-2026

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I Recall

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

FDA recall Class I

Ongoing

FDA lists this Class I recall for NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0029-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Sep 9, 2025
FDA report date
Oct 15, 2025
Quantity in commerce
1667 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

Affected model & UDI checker

Search the exact identifiers published for Z-0029-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedNOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-ICheck source notice05060541640009
FDA code information

Model Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.

View the original FDA recall notice

Why was this product recalled?

The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.

What actions did NOXBOX LTD take?

On September 9, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers: Actions to be taken: - Customers should ensure that all affected personnel are provided with a copy of the recommendations in this letter. - Users should avoid rapidly pressing buttons in quick succession (e.g., go to next screen, accept a value, change a value, backing out, repeating this) as it is necessary to allow an opportunity for the device to respond to user inputs before the next command to avoid a potential software interruption and subsequent restart. - NOxBOX Ltd. is asking customers in the U.S. to return the attached response form to Linde Gas & Equipment as instructed on the form. Recalling firm actions: NOxBOX Ltd. is in the process of developing a new software version that will include a modification to the software to resolve this issue, which will be released pending required regulatory review and approval. Once the software version is released, NOxBOX Ltd. will be back in contact with you to update your device software. If you have any questions about this notice or need assistance, please reach out to your local Clinical Educator or the Customer Excellence Center at 1-833-NOX-VENT (669-8368) or at LG.US.NOXIVENT@Linde.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.