DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer Recall
DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer
FDA recall Class II
Ongoing
FDA lists this Class II recall for DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0032-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 4, 2025
- FDA report date
- Oct 8, 2025
- Quantity in commerce
- 116
- Distribution
- See key details
- FDA cause
- Software Design Change
What is being recalled?
DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer
Affected model & UDI checker
Search the exact identifiers published for Z-0032-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| C63522 | DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer | Check source notice | Not specified |
| C6352 | DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer | Check source notice | Not specified |
FDA code information
REF/UDI-DI: C63522/14987666545089, C63521/14987666545072. Software Versions: SW 1.3.3 and 1.3.4
Why was this product recalled?
On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.
What actions did BECKMAN COULTER, INC. take?
On 8/4/2025, recall notices were mailed and/or emailed to customers who were asked to do the following: Check your software version, select Menu - About page. - For Clinical Analyzers running software version 1.3.3 or 1.3.4, if this issue occurs, you will need to place a new order for the CC combination test, - Select Home - Sample List task indicator, then select Active in the upper left corner of the Sample List to display samples "In Progress". - Program the CC combination tests for these "In Progress" samples using a new rack and Sample ID, apply the new barcode, then load the rack onto the analyzer for processing. - Set aside the rack with "In Progress" samples and contact your Beckman Coulter representative to clear them. Firm recommends sharing the content of this letter with your laboratory and/or Medical Director. Firm has resolved this issue in Software version 1.3.5 and a firm service representative will contact you to schedule the software upgrade. Share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. Complete and return the response form via email to: regaffairsfax@beckman.com If you have any questions regarding this product, please contact your local firm representative, or use the following link for a listing of local contact information. https://www.beckmancoulter.com/en/support/contact-us
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0032-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.