CorNeat EverPatch, REF: 1036295, Permanent Tissue-Integrating Surgical Matrix, 5.0 mm x 605 mm x 100 m Recall

CorNeat EverPatch, REF: 1036295, Permanent Tissue-Integrating Surgical Matrix, 5.0 mm x 605 mm x 100 ¿m

FDA recall Class II

Ongoing

FDA lists this Class II recall for CorNeat EverPatch, REF: 1036295, Permanent Tissue-Integrating Surgical Matrix, 5.0 mm x 605 mm x 100 m. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0039-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 18, 2025
FDA report date
Oct 15, 2025
Quantity in commerce
630 units
Distribution
See key details
FDA cause
Unknown/Undetermined by firm

What is being recalled?

CorNeat EverPatch, REF: 1036295, Permanent Tissue-Integrating Surgical Matrix, 5.0 mm x 605 mm x 100 m

Affected model & UDI checker

Search the exact identifiers published for Z-0039-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1036295CorNeat EverPatch, REF: 1036295, Permanent Tissue-Integrating Surgical Matrix, 5.0 mm x 605 mm x 100 mCheck source noticeNot specified
FDA code information

All Lots/ UDI: G16010362950

View the original FDA recall notice

Why was this product recalled?

Reviewed complaints have identified that conjunctival wound dehiscence may result in early exposure of the surgical patch.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did CORNEAT VISION, LTD. take?

On August 18, 2025, CorNeat Vision, Ltd. issued a "Urgent: Medical Device Recall" Notification to affected consignees via E-Mail. CorNeat ask consignees to take the following actions: 1. Review inventory, Quarantine and Discontinue use of All Lots of the CorNeat EverPatch and/or EverPatch+. 2. Please discard any unused product. 3. Eyecare providers should continue to monitor patients who have had CorNeat EverPatch and/or EverPatch+ implanted at the follow-up frequency necessary to detect signs of wound dehiscence or conjunctival thinning/erosion. 4. Patients should be instructed to seek prompt follow-up if they experience any signs or symptoms consistent with this complication. 5. This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 6. Please fill in the MEDICAL DEVICE RECALL RETURN RESPONSE form at www.corneat.com/everpatch-recall or fill in the attached form and forward to CorNeat Vision by email to complaints@corneat.com .

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology