EverPatch+, REF: 1044002, Permanent Scleral Reinforcement Matrix 1x5.0mm x 8.5 mm Shield/1 x 5.0mm x 6.5 mm Rectangle, STERILE VH202 Recall
EverPatch+, REF: 1044002, Permanent Scleral Reinforcement Matrix 1x5.0mm x 8.5 mm Shield/1 x 5.0mm x 6.5 mm Rectangle, STERILE VH202
FDA recall Class II
Ongoing
FDA lists this Class II recall for EverPatch+, REF: 1044002, Permanent Scleral Reinforcement Matrix 1x5.0mm x 8.5 mm Shield/1 x 5.0mm x 6.5 mm Rectangle, STERILE VH202. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0040-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 18, 2025
- FDA report date
- Oct 15, 2025
- Quantity in commerce
- 180 units
- Distribution
- See key details
- FDA cause
- Unknown/Undetermined by firm
What is being recalled?
EverPatch+, REF: 1044002, Permanent Scleral Reinforcement Matrix 1x5.0mm x 8.5 mm Shield/1 x 5.0mm x 6.5 mm Rectangle, STERILE VH202
Affected model & UDI checker
Search the exact identifiers published for Z-0040-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 1044002 | EverPatch+, REF: 1044002, Permanent Scleral Reinforcement Matrix 1x5.0mm x 8.5 mm Shield/1 x 5.0mm x 6.5 mm Rectangle, STERILE VH202 | Check source notice | Not specified |
FDA code information
All Lots/ UDI: G16010440020
Why was this product recalled?
Reviewed complaints have identified that conjunctival wound dehiscence may result in early exposure of the surgical patch.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did CORNEAT VISION, LTD. take?
On August 18, 2025, CorNeat Vision, Ltd. issued a "Urgent: Medical Device Recall" Notification to affected consignees via E-Mail. CorNeat ask consignees to take the following actions: 1. Review inventory, Quarantine and Discontinue use of All Lots of the CorNeat EverPatch and/or EverPatch+. 2. Please discard any unused product. 3. Eyecare providers should continue to monitor patients who have had CorNeat EverPatch and/or EverPatch+ implanted at the follow-up frequency necessary to detect signs of wound dehiscence or conjunctival thinning/erosion. 4. Patients should be instructed to seek prompt follow-up if they experience any signs or symptoms consistent with this complication. 5. This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 6. Please fill in the MEDICAL DEVICE RECALL RETURN RESPONSE form at www.corneat.com/everpatch-recall or fill in the attached form and forward to CorNeat Vision by email to complaints@corneat.com .
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0040-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.