FDA RECALL RECORD · Z-0070-2026

Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG Recall

Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG Model/Catalog Number: SL-2010M2096 Software Version: N/A Product Description: STREAMLINE BLOODLINE SET FOR DIALOG Component: N/A

FDA recall Class I

Ongoing

FDA lists this Class I recall for Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0070-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Sep 30, 2025
FDA report date
Oct 29, 2025
Quantity in commerce
1298454 units (1589674 - 3/5/26 Expansion)
Distribution
See key details
FDA cause
Process control

What is being recalled?

Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG Model/Catalog Number: SL-2010M2096 Software Version: N/A Product Description: STREAMLINE BLOODLINE SET FOR DIALOG Component: N/A

Affected model & UDI checker

Search the exact identifiers published for Z-0070-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
SL-2010M2096Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOGCheck source noticeNot specified
Not specifiedBrand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOGCheck source notice04046964367786
FDA code information

Model No. SL-2010M2096; UDI-DI 04046964367786; Lot Numbers: A2500119, A2500121, A2500128, A2500129, A2500130, A2500132, A2500133, A2500134, A2500149, A2500150, A2500151, A2500180, A2500181, A2500182, A2500185, A2500198, A2500199, A2500200, A2500203, A2500204, A2500205, A2500206, A2500207, A2500208, A2500209, A2500210, A2500211, A2500212, A2500213, A2500214, A2500215, A2500216, A2500217, A2500218, A2500219, A2500220, A2500246, A2500247, A2500248, A2500249, A2500250, A2500251, A2500257, A2500258, A2500259, A2500260, A2500261, A2500262, A2500267, A2500268, A2500269, A2500273, A2500274, A2500275, A2500276, A2500277, A2500278, A2500295, A2500298, A2500299, A2500300; ***Addded 3/24/2026***Additional Lots Identified A2500318, A2500319, A2500320, A2500321, A2500322, A2500323, A2500324, A2500325, A2500326, A2500327, A2500328, A2500329, A2500330, A2500356, A2500357;

View the original FDA recall notice

Why was this product recalled?

The potential for micro-air bubbles observed in the bloodline and air-in-line alarms due to damaged arterial and venous patient connectors.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did B. BRAUN MEDICAL INC. take?

On September 29, 2025, "URGENT MEDICAL DEVICE CORRECTION" letters dated September 24, 2025, were sent to customers. Actions to be taken: 1. Ensure full adherence to the product Instructions for Use (IFU). In particular, the instructions for secure connection of patient connectors should be followed to reduce the probability of air ingress or leakage. 2. Increase visual inspection of the bloodlines during treatment. If air ingress or leakage is observed, re-secure or tighten connection using the product IFU. In some instances, the dialysis treatment should be stopped. 3. If re-securement or tightening connections does not resolve the concern, replace bloodline at earliest clinically feasible time. 4. If possible, avoid high flow during treatment if microbubbles are present. B. Braun Medical Inc. (BBMI) Customer/User: 1. Review this notice in its entirety. Ensure that all users in your organization of the above-mentioned products, and other concerned personnel are informed about this voluntary correction. Post this notification where the affected products are stored. Note: Product is not being returned as part of this correction notification. a. If you are a distributor and have further distributed the product, please forward this notice to your consignees. The correction is to be extended to the consumer level. 2. Return the completed Urgent Medical Device Correction Acknowledgement Form to B. Braun Medical Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to recalls@bbraunusa.com within two (2) weeks of receipt, even if the total inventory in your possession is zero (0). If you have any questions, contact Director, Corporate Communications at 484-240-8373. ****Update 03/24/2026****On March 5, 2026, URGENT MEDICAL DEVICE CORRECTION letters were sent update letters out to notify of additional 15 lots impacted by the recall.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology