SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only. Recall
SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only.
FDA recall Class II
Ongoing
FDA lists this Class II recall for SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0071-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 3, 2025
- FDA report date
- Oct 15, 2025
- Quantity in commerce
- 2146 units
- Distribution
- See key details
- FDA cause
- Unknown/Undetermined by firm
What is being recalled?
SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only.
Affected model & UDI checker
Search the exact identifiers published for Z-0071-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 467868 | SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only. | Check source notice | Not specified |
| Not specified | SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only. | Check source notice | 15099590233358 |
FDA code information
UDI: 15099590233358/ Lot Numbers: M309642 M309641 M401490 M404365 M406380 M406417 M408408 M410466 M410513 M412278
Why was this product recalled?
Specific SYNCHRON Systems Modular Chemistry Phosphorus (PHOSm REF 467868) reagent lots may fail calibration before the lot expiration date which causes a delay in obtaining patient results.
What actions did BECKMAN COULTER, INC. take?
On September 3, 2025, Beckman Coulter issued a "Urgent Medical Device Recall" Notification to affected consignees via E-Mail. Beckman Coulter asked consignees to take the following actions: 1. You may continue to use the affected lots if calibration passes and quality control results are within your facility s acceptance criteria. 2. If the above lots fail calibration due to INIT ADC HI before the lot expiration date, contact Customer Technical Support Center for troubleshooting and/or replacement. 3. If you receive replacement product, review the Safety Data Sheet (SDS) and discard the affected lot(s). 4. Beckman Coulter offers SYNCHRON Systems Cartridge Chemistry Phosphorus (PHS REF A09426) Reagent as an alternative. 5. Please share this information with your laboratory staff. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. 6. Please respond within 10 days.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0071-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026