Rollator, Bariatric, Basic Steel, Burgundy, mechanical walker with four legs and a metal frame intended for medical purposes to provide moderate weight support while walking, Product Number MDS86800XWS Recall
Rollator, Bariatric, Basic Steel, Burgundy, mechanical walker with four legs and a metal frame intended for medical purposes to provide moderate weight support while walking, Product Number MDS86800XWS
FDA recall Class II
Ongoing
FDA lists this Class II recall for Rollator, Bariatric, Basic Steel, Burgundy, mechanical walker with four legs and a metal frame intended for medical purposes to provide moderate weight support while walking, Product Number MDS86800XWS. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0086-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 14, 2026
- FDA report date
- Oct 14, 2026
- Quantity in commerce
- 1824 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Rollator, Bariatric, Basic Steel, Burgundy, mechanical walker with four legs and a metal frame intended for medical purposes to provide moderate weight support while walking, Product Number MDS86800XWS
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Lot Numbers: 05425090001 05425120001 05426010001 05426020001 05426030001 05426040001 05426050001
Why was this product recalled?
The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from the rollator frame tubing, the front wheels may become loose or detach from the rollator. If the front wheels become loose or detach during use, there is risk of severe adverse events. The immediate consequences could include falls resulting in head trauma, bone fractures, or other critical injuries.
What actions did Medline Industries, LP take?
An URGENT MEDICAL DEVICE RECALL notification letter was sent to customers beginning 8/14/26. REQUIRED ACTION: 1. Inspect Your Rollator Immediately " Please refer to Image 1: How to locate and identify the product Lot Number to locate the front and rear side of the rollator. " On the rear side of the rollator, you will find the Lot Label is located on the left side frame tubing, below the seat frame tubing. 2. If your rollator matches both the affected SKU and an affected Lot Number, immediately discontinue use of the device. Take a clear photograph of the Lot Label and contact Medline Industries, LP directly, as soon as possible. A photograph of the Lot Number must be submitted to Medline for verification. You may contact Medline Industries, LP using one of the following methods: " Email: QualityAssurance@medline.com " Phone: 1-800-950-0128 option 5 Once Medline receives and confirms your Lot Number, further instructions regarding discontinued use of the affected rollator will be provided. 3. If your rollator does not have one of the affected Lot Numbers, it may continue to be used in accordance with the Instructions for Use. Please complete and return the attached response form only if you do not have the affected rollator Lot Number. 4. If you have resold, distributed, or transferred this product to another company or individual, FDA regulations require that you notify those recipients of this recall communication. Please contact Medline Industries, LP for further instructions regarding the identification, notification, and handling of affected products distributed to your customers. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0086-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology