FDA RECALL RECORD · Z-0089-2027

Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station Recall

Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station, Model/Catalog Number: CN-6201 Software Version: software versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40 Product Description: The device is intended for use by medical professionals to provide cardiac and vital signs monitoring for multiple patients within a medical facility. The CNS-6200 Series Central Nurse Station will display and record physiological data from up to forty telemetry receiver/transmitters and generates an alarm when a measured parameter falls outside a pre-set limit or when life threatening arrhythmia is detected. Arrhythmia detection and alarm determination are functions of the telemetry receivers/transmitters or individual bedside monitor. Component: No

FDA recall Class II

Ongoing

FDA lists this Class II recall for Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0089-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 8, 2026
FDA report date
Oct 14, 2026
Quantity in commerce
808
Distribution
See key details
FDA cause
Software design

What is being recalled?

Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station, Model/Catalog Number: CN-6201 Software Version: software versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40 Product Description: The device is intended for use by medical professionals to provide cardiac and vital signs monitoring for multiple patients within a medical facility. The CNS-6200 Series Central Nurse Station will display and record physiological data from up to forty telemetry receiver/transmitters and generates an alarm when a measured parameter falls outside a pre-set limit or when life threatening arrhythmia is detected. Arrhythmia detection and alarm determination are functions of the telemetry receivers/transmitters or individual bedside monitor. Component: No

Affected model & UDI checker

Search the exact identifiers published for Z-0089-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
CN-6201Product Name: Nihon Kohden CNS-6200 Series Central Nurse StationCheck source noticeNot specified
FDA code information

Lot Code: CNS-6201 (PU-621RA), SW versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40

View the original FDA recall notice

Why was this product recalled?

Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Nihon Kohden America LLC take?

On 09/08/2026, an "URGENT: MEDICAL DEVICE CORRECTION" Letter was sent via USPS certified mail to customers informing that, Nihon Kohden Corporation announced the discovery of a vulnerability within a limited number of CNS-6201 models to Denial of Service (DoS) attacks. The identified vulnerability is classified as a NULL pointer dereference* (CWE-476). This type of software error may cause the system to crash when processing malformed or unexpected network input. Nihon Kohden has not received any complaints related to this vulnerability. If a CNS-6201 is affected by a DoS event, the device may stop its monitoring function due to a system crash, resulting in temporary loss of central monitoring functionality. Customers are instructed to: 1. Identify any affected CNS-6201 units in their inventory. 2. Complete the attached Medical Device Notification Return Response Form and email the completed Medical Device Notification Return Response Form back to NKA Technical Services at vitalsignsTS@nihonkohden.com. Issue Resolution a. Nihon Kohden Corporation advises that the customer upgrade to a successor product (e.g. CNS-2101). b. For customers who continue to operate CNS-6201 systems, Nihon Kohden reiterates the importance of implementing established cybersecurity controls that are part of the intended deployment architecture and long-standing cybersecurity guidance for these systems. These controls are not newly introduced mitigations but rather represent expected deployment requirements designed to support the secure operation of the system throughout its lifecycle. For questions - call Nihon Kohden America at 1-800-325-0283, Option 2 (M-F, 8am-5pm PDT) or email vitalsignsTS@nihonkohden.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology