Onkos Surgical Inc. My3D Personalized Pelvic Implant, Porous, Left REF PV-ACREC-04M Recall
Onkos Surgical Inc. My3D Personalized Pelvic Implant, Porous, Left REF PV-ACREC-04M
FDA recall Class II
Ongoing
FDA lists this Class II recall for Onkos Surgical Inc. My3D Personalized Pelvic Implant, Porous, Left REF PV-ACREC-04M. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0090-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 24, 2026
- FDA report date
- Oct 14, 2026
- Quantity in commerce
- 1 implant
- Distribution
- See key details
- FDA cause
- Labeling mix-ups
What is being recalled?
Onkos Surgical Inc. My3D Personalized Pelvic Implant, Porous, Left REF PV-ACREC-04M
Affected model & UDI checker
Search the exact identifiers published for Z-0090-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| PV-ACREC-04M | Onkos Surgical Inc. My3D Personalized Pelvic Implant, Porous, Left REF PV-ACREC-04M | Check source notice | Not specified |
FDA code information
UDI: B278PVACREC04M0-Lot# C26-0346; Case# C26-0346,
Why was this product recalled?
Personalized pelvic implants inadvertently mixed up and distributed to incorrect customers.
What actions did Onkos Surgical, Inc. take?
On July 24, 2026, Onkos Surgical Inc. issued a medical device recall to affected consignees via email. Onkos Surgical Inc. informed consignees of the following actions to be taken: 1.Segregate and quarantine the product immediately. 2.Within two business days, complete the Product and Distribution Table within the medical device recall document and acknowledge receipt of this letter. 3. Return your existing implants for inspection within 2 business days after signing the acknowledgement. Return to: Onkos Surgical, Inc., Attention: Matt Avioli, 77 East Halsey Road, Parsippany, NJ 07054 4. Contact Matt Avioli for questions:Phone: 781-223-0938, Email contact: mavioli@onkossurgical.com
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0090-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.