Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0108-2027

Alphenix INFX-8000C/B5 Recall

Alphenix INFX-8000C/B5

FDA recall Class II

Ongoing

FDA lists this Class II recall for Alphenix INFX-8000C/B5. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0108-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 26, 2026
FDA report date
Oct 14, 2026
Quantity in commerce
1 unit
Distribution
See key details
FDA cause
Process control

What is being recalled?

Alphenix INFX-8000C/B5

Affected model & UDI checker

Search the exact identifiers published for Z-0108-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedAlphenix INFX-8000C/B5Check source notice4987670100192
FDA code information

Serial no. B5F2532011/UDI: 4987670100192

View the original FDA recall notice

Why was this product recalled?

Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error

What actions did Canon Medical System, USA, INC. take?

On August 26, 2026, Canon Medical Systems USA, Inc. issued a Urgent: Medical Device Correction notification to affected consignees via mail. Canon ask consignees to take the following actions: 1. Should this problem occur before inspection of the tightening of the bolt(s) is implemented on your system, please stop using the device and call your service representative. 2. Additionally, it is strongly requested that you share the contents of this letter with all users as well as clinical engineering or biomedical group at your facility. 3. A representative of Canon Medical Systems USA will contact you to schedule a time to inspect the tightening and properly fixing of the bolt(s) to correct this issue with your system(s). 4. If you have any questions regarding this matter, please contact your service representative contact our InTouch team at (800) 521-1968 or by email at intouchdispatch@us.medical.canon. 5. Canon respectfully requests that the attached form be completed and returned via email to fieldcorrections@us.medical.canon.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.