PERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (9cm) Tuohy Epidural Needle - Winged and PERIFIX FX Springwound Catheter - 19 Ga. Closed Tip. Product Code: CE17TBFC. Product description: CE17TBFC EPIDURAL BASIC TRAY 17G TOUHY. Recall
PERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (9cm) Tuohy Epidural Needle - Winged and PERIFIX FX Springwound Catheter - 19 Ga. Closed Tip. Product Code: CE17TBFC. Product description: CE17TBFC EPIDURAL BASIC TRAY 17G TOUHY.
FDA recall Class II
Ongoing
FDA lists this Class II recall for PERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (9cm) Tuohy Epidural Needle - Winged and PERIFIX FX Springwound Catheter - 19 Ga. Closed Tip. Product Code: CE17TBFC. Product description: CE17TBFC EPIDURAL BASIC TRAY 17G TOUHY.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0114-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 18, 2025
- FDA report date
- Oct 15, 2025
- Quantity in commerce
- 0 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
PERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (9cm) Tuohy Epidural Needle - Winged and PERIFIX FX Springwound Catheter - 19 Ga. Closed Tip. Product Code: CE17TBFC. Product description: CE17TBFC EPIDURAL BASIC TRAY 17G TOUHY.
Affected model & UDI checker
Search the exact identifiers published for Z-0114-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 332077 | PERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (9cm) Tuohy Epidural Needle - Winged and PERIFIX FX Springwound Catheter - 19 Ga. Closed Tip. Product Code: CE17TBFC. Product description: CE17TBFC EPIDURAL BASIC TRAY 17G TOUHY. | Check source notice | Not specified |
FDA code information
REF: 332077; Product Code: CE17TBFC; UDI-DI (Primary): 4046964177187; UDI-DI (Unit of Use): 4046964177170; Lot Number: 0062022053. Expiration Date: 10/16/2024
Why was this product recalled?
Potential for the lid of the catheter connector to be in the incorrect position.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did B. BRAUN MEDICAL INC. take?
B Braun notified consignees on about 09/18/2025 via USPS certified mailed URGENT MEDICAL DEVICE CORRECTION NOTIFICATION. Consignees were instructed to, where possible, have a second kit or additional catheter connector available at the point of use and upon opening the kit, verify the correct catheter connector is present. If the affected component is detected, discard and do not use. Consignees were also instructed to review the Medical Device Correction Notification and ensure all users in your organization and other concerned persons are informed, if affected units have been further distributed, notify customers, post the notification where affected units are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form." Additionally, if consignees have identified affected units in inventory, they may order one case (50 pieces) of affected sterilized single use catheter connectors to have on hand.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0114-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026