TufTex Over-the-Wire Embolectomy Catheter; Recall
TufTex¿ Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;
FDA recall Class II
Ongoing
FDA lists this Class II recall for TufTex Over-the-Wire Embolectomy Catheter;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0117-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 28, 2026
- FDA report date
- Oct 14, 2026
- Quantity in commerce
- 54
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
TufTex Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;
Affected model & UDI checker
Search the exact identifiers published for Z-0117-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| E1651-88 | TufTex Over-the-Wire Embolectomy Catheter; | Check source notice | Not specified |
| Not specified | TufTex Over-the-Wire Embolectomy Catheter; | Check source notice | 00840663100866 |
FDA code information
Catalog Number: E1651-88; UDI-DI: 00840663100866; Lot Number: XOT00002181;
Why was this product recalled?
Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.
What actions did LeMaitre Vascular, Inc. take?
On August 28, 2026 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: 1. Check your inventory against the list of lots in this letter. Immediately quarantine any recalled devices. 2. Complete the form at the end of this letter. Please note that you must return the form even if you have no devices in inventory. 3. LeMaitre customer service will provide an RMA number to coordinate the product return. 4. When a recalled device has been returned to LeMaitre Vascular, a replacement device will be provided. 5. If you have transferred devices to another facility, please forward a copy of this recall letter to them.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0117-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology