Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0118-2026

PERIFIX FX Continuous Epidural Anesthesia Tray, 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Fixed Wings, 19 Ga. Open tip PERIFIX FX Springwound Epidural Catheter. Product Code: CE17TKFSDT. Product Description: CE17TKFSDT EPID TRAY W/17GA FIXED WING. Recall

PERIFIX FX Continuous Epidural Anesthesia Tray, 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Fixed Wings, 19 Ga. Open tip PERIFIX FX Springwound Epidural Catheter. Product Code: CE17TKFSDT. Product Description: CE17TKFSDT EPID TRAY W/17GA FIXED WING.

FDA recall Class II

Ongoing

FDA lists this Class II recall for PERIFIX FX Continuous Epidural Anesthesia Tray, 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Fixed Wings, 19 Ga. Open tip PERIFIX FX Springwound Epidural Catheter. Product Code: CE17TKFSDT. Product Description: CE17TKFSDT EPID TRAY W/17GA FIXED WING.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0118-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 18, 2025
FDA report date
Oct 15, 2025
Quantity in commerce
0 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

PERIFIX FX Continuous Epidural Anesthesia Tray, 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Fixed Wings, 19 Ga. Open tip PERIFIX FX Springwound Epidural Catheter. Product Code: CE17TKFSDT. Product Description: CE17TKFSDT EPID TRAY W/17GA FIXED WING.

Affected model & UDI checker

Search the exact identifiers published for Z-0118-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
332093PERIFIX FX Continuous Epidural Anesthesia Tray, 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Fixed Wings, 19 Ga. Open tip PERIFIX FX Springwound Epidural Catheter. Product Code: CE17TKFSDT. Product Description: CE17TKFSDT EPID TRAY W/17GA FIXED WING.Check source noticeNot specified
FDA code information

REF: 332093; Product Code: CE17TKFSDT; UDI-DI (Primary): 04046964954252; UDI-DI (Unit of Use): 04046964954245; Lot Number: 0062003112. Expiration Date: 07/31/2026.

View the original FDA recall notice

Why was this product recalled?

Potential for the lid of the catheter connector to be in the incorrect position.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did B. BRAUN MEDICAL INC. take?

B Braun notified consignees on about 09/18/2025 via USPS certified mailed URGENT MEDICAL DEVICE CORRECTION NOTIFICATION. Consignees were instructed to, where possible, have a second kit or additional catheter connector available at the point of use and upon opening the kit, verify the correct catheter connector is present. If the affected component is detected, discard and do not use. Consignees were also instructed to review the Medical Device Correction Notification and ensure all users in your organization and other concerned persons are informed, if affected units have been further distributed, notify customers, post the notification where affected units are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form." Additionally, if consignees have identified affected units in inventory, they may order one case (50 pieces) of affected sterilized single use catheter connectors to have on hand.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.