NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 6 F x 55, REF E NRT-05055-I The NutraGlide Nasal Feeding Tube is a feeding tube placed through the nose and ending in the stomach or intestine used for the administration of nutrition, fluids, and medications in neonatal, pediatric, and adult patients. Recall
NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 6 F x 55, REF E NRT-05055-I The NutraGlide Nasal Feeding Tube is a feeding tube placed through the nose and ending in the stomach or intestine used for the administration of nutrition, fluids, and medications in neonatal, pediatric, and adult patients.
FDA recall Class II
Completed
FDA lists this Class II recall for NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 6 F x 55, REF E NRT-05055-I The NutraGlide Nasal Feeding Tube is a feeding tube placed through the nose and ending in the stomach or intestine used for the administration of nutrition, fluids, and medications in neonatal, pediatric, and adult patients.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0127-2026
- Recall classification
- Class II
- Enforcement status
- Completed
- Initiation date
- Aug 12, 2025
- FDA report date
- Oct 15, 2025
- Quantity in commerce
- 1780 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 6 F x 55, REF E NRT-05055-I The NutraGlide Nasal Feeding Tube is a feeding tube placed through the nose and ending in the stomach or intestine used for the administration of nutrition, fluids, and medications in neonatal, pediatric, and adult patients.
Affected model & UDI checker
Search the exact identifiers published for Z-0127-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 6 F x 55, REF E NRT-05055-I The NutraGlide Nasal Feeding Tube is a feeding tube placed through the nose and ending in the stomach or intestine used for the administration of nutrition, fluids, and medications in neonatal, pediatric, and adult patients. | Check source notice | 00842071149628 |
| Not specified | NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 6 F x 55, REF E NRT-05055-I The NutraGlide Nasal Feeding Tube is a feeding tube placed through the nose and ending in the stomach or intestine used for the administration of nutrition, fluids, and medications in neonatal, pediatric, and adult patients. | Check source notice | 00842071149505 |
FDA code information
Box UDI (01)00842071149628(17)280601(10)250711-237, Pouch UDI: (01)00842071149505(17)280701(10)250707-412, (01)00842071149505(17)280601(10)250701-300. Box UDI (01)00842071149628(17)280701(10)250717-425, Pouch UDI: (01)00842071149505(17)280701(10)250710-247, (01)00842071149505(17)280701(10)250710-246. Box UDI (01)00842071149628(17)280601(10)250709-162, Pouch UDI: (01)00842071149505(17)280701(10)250702-222, (01)00842071149505(17)280601(10)250627-420, (01)00842071149505(17)280701(10)250702-412. Box UDI (01)00842071149628(17)280601(10)250710-217, Pouch UDI: (01)00842071149505(17)280701(10)250702-169, (01)00842071149505(17)280701(10)250702-138, (01)00842071149505(17)280701(10)250702-220, (01)00842071149505(17)280601(10)250630-093, (01)00842071149505(17)280601(10)250630-048, (01)00842071149505(17)280601(10)250630-047, (01)00842071149505(17)280601(10)250630-046, (01)00842071149505(17)280701(10)250702-170. Box UDI (01)00842071149628(17)280601(10)250710-307, Pouch UDI: (01)00842071149505(17)280701(10)250707-457, (01)00842071149505(17)280701(10)250702-414, (01)00842071149505(17)280601(10)250701-302, (01)00842071149505(17)280601(10)250701-301, (01)00842071149505(17)280701(10)250702-137. Box UDI (01)00842071149628(17)280601(10)250620-347, Pouch UDI: (01)00842071149505(17)280601(10)250617-066, (01)00842071149505(17)280601(10)250617-067, (01)00842071149505(17)280601(10)250617-101. Box UDI (01)00842071149628(17)280601(10)250626-400, Pouch UDI: (01)00842071149505(17)280601(10)250609-364. Box UDI (01)00842071149628(17)280601(10)250618-493, Pouch UDI: (01)00842071149505(17)280601(10)250609-360, (01)00842071149505(17)280601(10)250604-076, (01)00842071149505(17)280601(10)250611-223, (01)00842071149505(17)280601(10)250612-097, (01)00842071149505(17)280601(10)250609-363, (01)00842071149505(17)280601(10)250609-362. Box UDI (01)00842071149628(17)280501(10)250612-388, Pouch UDI: (01)00842071149505(17)280601(10)250605-139, (01)00842071149505(17)280601(10)250603-078, (01)00842071149505(17)280601(10)250604-383. Box UDI (01)00842071149628(17)280601(10)250616-372, Pouch UDI: (01)00842071149505(17)280601(10)250604-075, (01)00842071149505(17)280601(10)250604-382, (01)00842071149505(17)280601(10)250605-338.
Why was this product recalled?
The nasal feeding tubes, NutraGlide, subject to this recall have distal tips that may detach at lower than expected forces.
What actions did Applied Medical Technology Inc take?
AMT issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 8/12/2025 via USPS certified mail. The notice explained the issue, risk to health and requested the following: ACTIONS TO BE TAKEN: CUSTOMER OR USER: Applied Medical Technology, Inc. requests that customers IMMEDIATELY TAKE THE FOLLOWING ACTIONS: " Examine your inventory and isolate/quarantine pending return of the product subject to this recall. " Do not use any device subject to this recall. " Complete and return to Applied Medical Technology, Inc. the enclosed RECALL NOTIFICATION ACKNOWLEDGMENT via fax or send by mail. " Please contact Applied Medical Technology, Inc. either by phone at 440-717-4000, or via email at cs@appliedmedical.net, to make arrangements for the return of any devices related to this recall. Include the product, all of its components, and any available packaging and labeling associated with the device.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0127-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology