Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0131-2027

VC(Japan) Recall

VC(Japan) Model Number: PRT-01098-100, Package PN:PRT-01125-100. used for ventilatory support.

FDA recall Class I

Ongoing

FDA lists this Class I recall for VC(Japan). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0131-2027
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Sep 17, 2026
FDA report date
Oct 28, 2026
Quantity in commerce
180 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

VC(Japan) Model Number: PRT-01098-100, Package PN:PRT-01125-100. used for ventilatory support.

Affected model & UDI checker

Search the exact identifiers published for Z-0131-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedVC(Japan)Check source notice00850018761116
FDA code information

All Serial Numbers: UDI: 00850018761116

View the original FDA recall notice

Why was this product recalled?

Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Ventec Life Systems, Inc. take?

On September 17, 2026, Ventec Life Systems Inc. dba React Health issued a Urgent: Medical Device Recall Notification via certified mail. Ventec ask consignees to take the following actions: 1.Review the customer communication and follow the applicable Actions to be Taken section. 2. Confirm receipt of this notification letter by returning the enclosed Customer Response Form to React Health via email, mail or use the online customer response form within 7 days. 3. If you have distributed or transferred affected devices to other facilities, caregivers, or end users, you must promptly notify them and provide a copy of this notification.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology