FERNO POWER X2 POWERED AMBULANCE COT Recall
FERNO POWER X2 POWERED AMBULANCE COT, Model/Catalog Number: 0015816; FOR USE IN GROUND TRANSPORTATION AMBULANCES.
FDA recall Class II
Ongoing
FDA lists this Class II recall for FERNO POWER X2 POWERED AMBULANCE COT. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0133-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 11, 2025
- FDA report date
- Oct 22, 2025
- Quantity in commerce
- 138 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
FERNO POWER X2 POWERED AMBULANCE COT, Model/Catalog Number: 0015816; FOR USE IN GROUND TRANSPORTATION AMBULANCES.
Affected model & UDI checker
Search the exact identifiers published for Z-0133-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 0015816 | FERNO POWER X2 POWERED AMBULANCE COT | Check source notice | Not specified |
| Not specified | FERNO POWER X2 POWERED AMBULANCE COT | Check source notice | 00190790003210 |
FDA code information
UDI-DI 00190790003210, Serial Numbers: 25G036127,25G036128, 25G036151, 25G036152, 25G036153, 25G036154, 25G036155, 25G036156, 25G036157, 25G036158, 25G036267, 25G036268, 25G036269, 25G036612, 25G036613, 25G036614, 25G036652, 25G036653, 25G036654, 25G036801, 25G036802, 25G036803, 25G036804, 25G036827, 25G036828, 25G036829, 25G036830, 25G036833, 25G036834, 25G036835, 25G036836, 25G036837, 25G036921, 25G036922, 25G036923, 25G036924, 25G036925, 25G036926, 25G036927, 25G036928, 25G036929, 25G036930, 25G036951, 25G036952, 25G036953, 25G036954, 25G036958, 25G036959, 25G036960, 25G036961, 25G036962, 25G036963, 25G036964, 25G037067, 25G037068, 25G037069, 25G037070, 25G037071, 25G037072, 25G037093, 25G037094, 25G037095, 25G037096, 25G037097, 25G037098, 25G037100, 25G037159, 25G037160, 25G037161, 25G037162, 25G037163, 25G037165, 25G037215, 25G037216, 25G037217, 25G037218, 25G037219, 25G037222, 25G037223, 25G037224, 25G037225, 25G037226, 25G037227, 25G037228, 25G037229, 25G037271, 25G037272, 25G037273, 25G037274, 25G037275, 25G037276, 25G037277, 25G037278, 25G037279, 25G037280, 25G037281, 25G037282, 25G037283, 25G037284, 25G037285, 25G037286, 25G037287, 25G037288, 25G037289, 25G037298, 25G037299, 25G037300, 25G037301, 25G037302, 25G037303, 25G037304, 25G037305, 25G037336, 25G037337, 25G037338, 25G037339, 25G037340, 25G037341, 25G037342, 25G037344, 25G037345, 25G037346, 25G037347, 25G037466, 25G037467, 25G037468, 25G037469, 25G037470, 25G037471, 25G037472, 25G037473, 25G037474, 25G037526, 25G037595, 25G037596, 25G037597, 25G037598, 25G037599, 25G037600, 25G037601, 25G037602, 25G037623, 25G037665, 25G037691, 25G037692, 25G037703, 25G037704, 25G037707, 25G037766, 25G037767, 25G037768, 25G037769, 25G037770, 25G037772, 25G037774, 25G037780, 25G037781, 25G037782, 25G037786, 25G038074, 25G038075, 25G038076, 25G038077, 25G038079, 25G038082, 25G038083, 25G038084, 25G038088, 25G038089, 25G038090, 25G038536, 25G038537, 25G038538, 25G038539, 25G038540, 25G038564, 25G038575, 25G038576, 25G038577, 25G038578, 25G038615, 25G038616, 25G038617, 25G038618, 25G038619, 25G038625, 25G038626, 25G038627, 25G038628, 25G038698, 25G038754, 25G038765, 25G038766, 25G038767, 25G038768, 25G038769
Why was this product recalled?
The fastening post hardware could loosen allowing the fastening post to detach from the cot
What actions did Ferno-Washington Inc take?
FERNO issued an URGENT PRODUCT CORRECTION notice to its consignees on 09/11/2025 via USPS certified mail. The notice explained the issue, potential risk to user, and requested the consignee identify the affected unit, inspect it using the following criteria: Visual Inspection " Look for gaps between the fastener head and the surface. " Check for missing fasteners. " Look for shiny wear marks, rust streaks, or metal shavings around the fastener these can indicate movement. " Inspect for vibration marks (circular or elongated wear around bolt holes). Physical/Manual Check " Hand-check: Use your fingers or a small wrench to see if the fastener turns easily when it should not. " Shake or tap test: Lightly tap the fastener or surrounding component with a non-marring tool; rattling or movement suggests looseness. Functional Signs " Listen for unusual rattling or vibration during operation. " Check for misalignment of connected parts (hinges, brackets, covers, panels). " Watch for loosened thread-locking compounds if broken, the fastener may have shifted. IF A COT IS FOUND TO HAVE LOOSE OR MISSING HARDWARE, YOU MUST IMMEDIATELY REMOVE IT FROM SERVICE UNTIL THE REPAIR IS COMPLETE. Upon receipt of this Notice, please call Ferno Technical Support at 866-987-3776 or email tscoordinator@ferno.com to (1) acknowledge receipt of this notice, (2) advise if any of your Power X2 serial numbers are affected, (3) advise if and how many of your affected serial numbers contain loose and/or missing hardware on the fastening post, and (4) make arrangements for trained service technicians to correct your ambulance cot. In the event you sold or donated the identified product, please forward this notice to your consignee and advise Ferno of the consignee information so we may follow up. Malfunctions or quality problems experienced with the use of your POWER X2 Powered Ambulance Cot may be reported to Ferno at 866-987-3776 or tscoordinator@ferno.com
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0133-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology