Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0141-2026

LGX Preconnect with TENACIO Pump without InhibiZone; UPNs: (1) 72404405 (12CM IP), (2) 72404407 (18CM IP); Recall

LGX Preconnect with TENACIO Pump without InhibiZone; UPNs: (1) 72404405 (12CM IP), (2) 72404407 (18CM IP);

FDA recall Class II

Ongoing

FDA lists this Class II recall for LGX Preconnect with TENACIO Pump without InhibiZone; UPNs: (1) 72404405 (12CM IP), (2) 72404407 (18CM IP);. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0141-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 11, 2025
FDA report date
Oct 22, 2025
Quantity in commerce
0 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

LGX Preconnect with TENACIO Pump without InhibiZone; UPNs: (1) 72404405 (12CM IP), (2) 72404407 (18CM IP);

Affected model & UDI checker

Search the exact identifiers published for Z-0141-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedLGX Preconnect with TENACIO Pump without InhibiZone; UPNs: (1) 72404405 (12CM IP), (2) 72404407 (18CM IP);Check source notice00191506021252
Not specifiedLGX Preconnect with TENACIO Pump without InhibiZone; UPNs: (1) 72404405 (12CM IP), (2) 72404407 (18CM IP);Check source notice00191506021276
FDA code information

1. UPN: 72404405 (12CM IP); UDI-DI: 00191506021252; Lot numbers: 1100687060, 1100687061; 2. UPN: 72404407 (18CM IP); UDI-DI: 00191506021276; Lot numbers: 1100676894, 1100676895;

View the original FDA recall notice

Why was this product recalled?

The potential for devices to experience inflation and/or deflation performance issues or difficulties.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did BOSTON SCIENTIFIC CORPORATION take?

On September 11, 2025 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: For implanted devices: If the device is working properly, no action is required. Further evaluation is recommended for patients who were implanted with an affected device AND are having difficulty inflating or deflating the device. - Check inventory for affected product and return any unused devices. Immediately cease further distribution or use of any unused affected devices. Please pass this notice to any healthcare professional from your organization that needs to be aware and to any organization where the potentially affected devices have been transferred. Please provide Boston Scientific with details of any affected devices that have been transferred to another organization.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.