Brand Name: ARVIS Product Name: ARVIS" Shoulder Recall
Brand Name: ARVIS Product Name: ARVIS" Shoulder Model/Catalog Number: IN-27300, Arvis Surgical Planning Software Software Version: V2025.1.2 Product Description: Intended Use: The ARVIS" system is a computer-controlled navigation system intended to provide intra-operative measurements to the surgeon to aid in selection and positioning of orthopedic implant components. Description: ARVIS Shoulder is a computer-controlled surgical navigation system for shoulder arthroplasty. It aids the surgeon in making intra-operative measurements and locating anatomical structures of the shoulder joint based on the patient s preoperative imaging to assist with selection and positioning of orthopedic implant components. The system consists of software, electronic hardware and surgical instruments. The device s workflow involves CT based preoperative planning followed by intraoperative navigation and execution. The preoperative planning software enables 3D virtual implant positioning based on the patient s CT reconstructed digital bone model and bony landmarks. The shoulder navigation application software then matches the patient s digital bone model and landmarks to the intraoperative landmarks registered by the surgeon, allowing an image-based navigation to follow. The surgeon uses the plan data as guidance to navigate and help position shoulder instruments and implants.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: ARVIS Product Name: ARVIS" Shoulder. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0145-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 19, 2025
- FDA report date
- Oct 22, 2025
- Quantity in commerce
- 21 systems
- Distribution
- See key details
- FDA cause
- Software change control
What is being recalled?
Brand Name: ARVIS Product Name: ARVIS" Shoulder Model/Catalog Number: IN-27300, Arvis Surgical Planning Software Software Version: V2025.1.2 Product Description: Intended Use: The ARVIS" system is a computer-controlled navigation system intended to provide intra-operative measurements to the surgeon to aid in selection and positioning of orthopedic implant components. Description: ARVIS Shoulder is a computer-controlled surgical navigation system for shoulder arthroplasty. It aids the surgeon in making intra-operative measurements and locating anatomical structures of the shoulder joint based on the patient s preoperative imaging to assist with selection and positioning of orthopedic implant components. The system consists of software, electronic hardware and surgical instruments. The device s workflow involves CT based preoperative planning followed by intraoperative navigation and execution. The preoperative planning software enables 3D virtual implant positioning based on the patient s CT reconstructed digital bone model and bony landmarks. The shoulder navigation application software then matches the patient s digital bone model and landmarks to the intraoperative landmarks registered by the surgeon, allowing an image-based navigation to follow. The surgeon uses the plan data as guidance to navigate and help position shoulder instruments and implants.
Affected model & UDI checker
Search the exact identifiers published for Z-0145-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| IN-27300 | Brand Name: ARVIS Product Name: ARVIS" Shoulder | Check source notice | Not specified |
FDA code information
Software Version V2025.1.2
Why was this product recalled?
Complaint identified issue with AI surgical planning software that may result in implant malalignment and/or decrease range of motion.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Kico Knee Innovation Company take?
On September 12, 2025, Kico Knee Innovation Company issued a recall notice to affected consignees. Kico asked consignees to take the following actions: 1. No actions are required by the surgeon users. 2. For the plans affected, Kico has quarantined these as they are not suitable for clinical use. 3. A long-term solution has been implemented to return the system to the specification of the most recently cleared device. We will contact surgeon users to login and re-plan their cases for upcoming surgeries.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0145-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology