BioPro Endo Head (Metal uni-polar, 38-55) Product Number/Description 10179 ENDO MD 38 10180 ENDO MD 41 10181 ENDO MD 43 10182 ENDO MD 45 10183 ENDO MD 47 10184 ENDO MD 49 10185 ENDO MD 51 10186 ENDO MD 53 10187 ENDO MD 55 10188 ENDO SH 38 10189 ENDO SH 41 10190 ENDO SH 43 10191 ENDO SH 45 10192 ENDO SH 47 10193 ENDO SH 49 10194 ENDO SH 51 10195 ENDO SH 53 10196 ENDO SH 55 Recall
BioPro Endo Head (Metal uni-polar, 38-55) Product Number/Description 10179 ENDO MD 38 10180 ENDO MD 41 10181 ENDO MD 43 10182 ENDO MD 45 10183 ENDO MD 47 10184 ENDO MD 49 10185 ENDO MD 51 10186 ENDO MD 53 10187 ENDO MD 55 10188 ENDO SH 38 10189 ENDO SH 41 10190 ENDO SH 43 10191 ENDO SH 45 10192 ENDO SH 47 10193 ENDO SH 49 10194 ENDO SH 51 10195 ENDO SH 53 10196 ENDO SH 55
FDA recall Class II
Ongoing
FDA lists this Class II recall for BioPro Endo Head (Metal uni-polar, 38-55) Product Number/Description 10179 ENDO MD 38 10180 ENDO MD 41 10181 ENDO MD 43 10182 ENDO MD 45 10183 ENDO MD 47 10184 ENDO MD 49 10185 ENDO MD 51 10186 ENDO MD 53 10187 ENDO MD 55 10188 ENDO SH 38 10189 ENDO SH 41 10190 ENDO SH 43 10191 ENDO SH 45 10192 ENDO SH 47 10193 ENDO SH 49 10194 ENDO SH 51 10195 ENDO SH 53 10196 ENDO SH 55. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0149-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 9, 2025
- FDA report date
- Oct 22, 2025
- Quantity in commerce
- 147 total
- Distribution
- See key details
- FDA cause
- Package design/selection
What is being recalled?
BioPro Endo Head (Metal uni-polar, 38-55) Product Number/Description 10179 ENDO MD 38 10180 ENDO MD 41 10181 ENDO MD 43 10182 ENDO MD 45 10183 ENDO MD 47 10184 ENDO MD 49 10185 ENDO MD 51 10186 ENDO MD 53 10187 ENDO MD 55 10188 ENDO SH 38 10189 ENDO SH 41 10190 ENDO SH 43 10191 ENDO SH 45 10192 ENDO SH 47 10193 ENDO SH 49 10194 ENDO SH 51 10195 ENDO SH 53 10196 ENDO SH 55
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Product ID/UDI-DI 10179 M20910179 10180 M20910180 10181 M20910181 10182 M20910182 10183 M20910183 10184 M20910184 10185 M20910185 10186 M20910186 10187 M20910187 10188 M20910188 10189 M20910189 10190 M20910190 10191 M20910191 10192 M20910192 10193 M20910193 10194 M20910194 10195 M20910195 10196 M20910196 All lots, all serial numbers
Why was this product recalled?
Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did BIOPRO, INC. take?
Consignees were notified on September 9, 2025, via phone, followed by formal letters with instructions for returning the inventory to BioPro.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0149-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.