Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0156-2026

LOGIQ P10 R4.5 HD ultrasound system with system software version R4.5.7., Model Number 5877535 Recall

LOGIQ P10 R4.5 HD ultrasound system with system software version R4.5.7., Model Number 5877535

FDA recall Class II

Ongoing

FDA lists this Class II recall for LOGIQ P10 R4.5 HD ultrasound system with system software version R4.5.7., Model Number 5877535. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0156-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 18, 2025
FDA report date
Oct 22, 2025
Quantity in commerce
32 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

LOGIQ P10 R4.5 HD ultrasound system with system software version R4.5.7., Model Number 5877535

Affected model & UDI checker

Search the exact identifiers published for Z-0156-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
5877535LOGIQ P10 R4.5 HD ultrasound system with system software version R4.5.7., Model Number 5877535Check source noticeNot specified
Not specifiedLOGIQ P10 R4.5 HD ultrasound system with system software version R4.5.7., Model Number 5877535Check source notice00195278464668
FDA code information

Model Number 5877535 UDI-DI: 00195278464668, Serial Numbers: LPZ490338 LPZ490347 LPZ490339 LPZ490147 LPZ490292 LPZ490019 LPZ490252 LPZ490348 LPZ490303 LPZ490304 LPZ490305 LPZ490306 LPZ490340 LPZ490350 LPZ490341 LPZ440658 LPZ440673 LPZ440674 LPZ490337 LPZ440641 LPZ440192 LPZ440258 LPZ440694 LPZ440695 LPZ440692 LPZ440632 LPZ440185 LPZ440107 LPZ440168 LPZ440046 LPZ440549

View the original FDA recall notice

Why was this product recalled?

The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did GE Medical Systems, LLC take?

An URGENT MEDICAL DEVICE CORRECTION notification letter dated 9/18/25 was sent to customers. Actions to be taken by Customer/User For LOGIQ P9 and LOGIQ P10 series ultrasound systems with system software version R4.5.7, discontinue using the UGAP feature (please see Appendix for product identification). You may continue using other imaging features. After the correction has been implemented, please destroy the installation media for software R4.5.7 at your site. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.74084@gehealthcare.com. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology