IMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida Recall
IMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida
FDA recall Class I
Ongoing
FDA lists this Class I recall for IMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0164-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Sep 30, 2025
- FDA report date
- Nov 5, 2025
- Quantity in commerce
- N/A
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
IMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida
Affected model & UDI checker
Search the exact identifiers published for Z-0164-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | IMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida | Check source notice | 00857534006806 |
FDA code information
IMRIS Part Number(s): 120823-000, UDI/DI 00857534006806
Why was this product recalled?
There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did IMPRIS Imaging Inc take?
Imris issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 9/30/2025 via email. The notice explained the issue, the risk to health, and provided the following information: "If the error message, "Magnet Supervision: The average heater power is above the alarm threshold. Please contact Siemens Service." appears, DO NOT REBOOT THE SYSTEM. Please call your local Siemens Healthineers service organization. Grant access to the system so that the inspection can be carried out as directed by Siemens Healthineers service organization." "Please immediately instruct your personnel accordingly, and please contact our customer support team via our service line at 1-866-475-0525 or e-mail at customersupport@imris.com if you have any questions or concerns. Please keep this information until the corrective measures have been finalized." This notice included a copy of the Siemens Healthineers Urgent Medical Device Correction notice.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0164-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.