Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135 LAP CHOLE DYNJ911660 LAP NISSEN DYNJ911560 UROLOGY LAPAROSCOPIC - 51638 DYNJ908720A Recall
Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135 LAP CHOLE DYNJ911660 LAP NISSEN DYNJ911560 UROLOGY LAPAROSCOPIC - 51638 DYNJ908720A
FDA recall Class II
Ongoing
FDA lists this Class II recall for Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135 LAP CHOLE DYNJ911660 LAP NISSEN DYNJ911560 UROLOGY LAPAROSCOPIC - 51638 DYNJ908720A. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0180-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 31, 2026
- FDA report date
- Oct 21, 2026
- Quantity in commerce
- 3,551 kits
- Distribution
- See key details
- FDA cause
- Process change control
What is being recalled?
Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135 LAP CHOLE DYNJ911660 LAP NISSEN DYNJ911560 UROLOGY LAPAROSCOPIC - 51638 DYNJ908720A
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
DYNJ911135 UDI-DI 10198459264382 Lots 25EDA279 25FDB612 25LDB658 DYNJ911660 UDI-DI 10198459487071 Lots 25IMF444 25JMA458 26BMG519 26DMA914 26EMA779 26FME775 26GMA849 26GMF826 26GMM101 26GMM127 DYNJ911560 UDI-DI 10198459400261 Lots 25HMG749 25JMC886 25KMA383 25KMA392 DYNJ908720A UDI-DI 10195327423155 Lots 24DBN292 24GBJ114 24IBK504 24JBC471 25KBP733 26ABG732
Why was this product recalled?
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Medline Industries, LP take?
The firm began notifying customers on August 31, 2026. Customers were provided with CONMED's original recall notification, which instructs users to have a replacement AirSeal disposable be available prior to the procedure in order to minimize potential delays. Users should ensure tubing is properly connected between the tube set and access ports prior to initiating insufflation. If the connection cannot be securely tightened and the "overpressure" alert cannot be resolved, users should replace the affected device. Users were also provided with instructions on how to connect the AirSeal tubeset and access ports, which include new figures as an update to the prior Instructions for Use. The corrected Instructions for Use are available electronically at eifu.conmed.com/eifu.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0180-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026