Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0180-2027

Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135 LAP CHOLE DYNJ911660 LAP NISSEN DYNJ911560 UROLOGY LAPAROSCOPIC - 51638 DYNJ908720A Recall

Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135 LAP CHOLE DYNJ911660 LAP NISSEN DYNJ911560 UROLOGY LAPAROSCOPIC - 51638 DYNJ908720A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135 LAP CHOLE DYNJ911660 LAP NISSEN DYNJ911560 UROLOGY LAPAROSCOPIC - 51638 DYNJ908720A. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0180-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 31, 2026
FDA report date
Oct 21, 2026
Quantity in commerce
3,551 kits
Distribution
See key details
FDA cause
Process change control

What is being recalled?

Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135 LAP CHOLE DYNJ911660 LAP NISSEN DYNJ911560 UROLOGY LAPAROSCOPIC - 51638 DYNJ908720A

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

DYNJ911135 UDI-DI 10198459264382 Lots 25EDA279 25FDB612 25LDB658 DYNJ911660 UDI-DI 10198459487071 Lots 25IMF444 25JMA458 26BMG519 26DMA914 26EMA779 26FME775 26GMA849 26GMF826 26GMM101 26GMM127 DYNJ911560 UDI-DI 10198459400261 Lots 25HMG749 25JMC886 25KMA383 25KMA392 DYNJ908720A UDI-DI 10195327423155 Lots 24DBN292 24GBJ114 24IBK504 24JBC471 25KBP733 26ABG732

View the original FDA recall notice

Why was this product recalled?

Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medline Industries, LP take?

The firm began notifying customers on August 31, 2026. Customers were provided with CONMED's original recall notification, which instructs users to have a replacement AirSeal disposable be available prior to the procedure in order to minimize potential delays. Users should ensure tubing is properly connected between the tube set and access ports prior to initiating insufflation. If the connection cannot be securely tightened and the "overpressure" alert cannot be resolved, users should replace the affected device. Users were also provided with instructions on how to connect the AirSeal tubeset and access ports, which include new figures as an update to the prior Instructions for Use. The corrected Instructions for Use are available electronically at eifu.conmed.com/eifu.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.