Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast Recall
Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast Model/Catalog Number: MEGABT09180, MEGABT09190, MEGABT091100, MEGABT091110 Software Version: N/A Product Description: The Mega Ballast 091 is designed with a semi-rigid proximal shaft and a highly flexible distal shaft. The design facilitates the advancement of the Mega Ballast to access the tortuous vasculature over a guidewire and a respective dilator. The outer surface of the Mega Ballast is coated with a hydrophilic coating to facilitate lubricity from tip to 40 cm proximal. The proximal end of the Mega Ballast incorporates a Luer fitting hub for the attachment of accessories. The Mega Ballast is provided sterile, non-pyrogenic, and is intended for single use only. Component: Not a component
FDA recall Class I
Ongoing
FDA lists this Class I recall for Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0188-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Sep 25, 2025
- FDA report date
- Nov 5, 2025
- Quantity in commerce
- 207 catheters
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast Model/Catalog Number: MEGABT09180, MEGABT09190, MEGABT091100, MEGABT091110 Software Version: N/A Product Description: The Mega Ballast 091 is designed with a semi-rigid proximal shaft and a highly flexible distal shaft. The design facilitates the advancement of the Mega Ballast to access the tortuous vasculature over a guidewire and a respective dilator. The outer surface of the Mega Ballast is coated with a hydrophilic coating to facilitate lubricity from tip to 40 cm proximal. The proximal end of the Mega Ballast incorporates a Luer fitting hub for the attachment of accessories. The Mega Ballast is provided sterile, non-pyrogenic, and is intended for single use only. Component: Not a component
Affected model & UDI checker
Search the exact identifiers published for Z-0188-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MEGABT09180 | Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast | Check source notice | Not specified |
| MEGABT09190 | Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast | Check source notice | Not specified |
| MEGABT091100 | Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast | Check source notice | Not specified |
| MEGABT091110 | Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast | Check source notice | Not specified |
FDA code information
Model Numbers and UDI-DI codes: Model No MEGABT09180 (00840303711780) Model No MEGABT09190 (00840303711797) Model No MEGABT091100 (00840303711803) Model No MEGABT091110 (00840303711810) Lot Numbers: F250600687, F250600686, F250600685, F250600455, F250600346, F250600345, F250600010, F250501434, F250501346, F250501091, F250500824, F250400219, F250400024, F250301020, F250300297, F250100956
Why was this product recalled?
Due to an identified manufacturing defect involving coating material lifting within the catheter's distal inner diameter/lumen.
What actions did Balt USA, LLC take?
On 09/18/2025, the firm sent via Priority Mail an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that, Balt USA has became aware of a situation where multiple Mega Ballast units failed the in-house QC final inspection due to the exterior coating material lifting. Further assessment found this coating also went inside of the ID and protruded from the inner surface of the catheter, which is not intended as part of the manufacturing process. Hospitals are instructed to: - Cease use of the affected product. - Inform hospital staff, such as safety officers, pharmacists, purchasing agent, head of neuroradiology and the neuroradiology staff, physicians including but not limited to endovascular neurosurgeons and interventional neurologists, as well as any other person if deemed necessary. - Identify and locate the Mega Ballast products concerned by this recall procedure. - Collect and put in quarantine the Mega Ballast products concerned by this recall, then return them to Balt USA through the usual administrative returns procedure by contacting Balt USA s Customer Service department. Distributors are instructed to: - Cease processing of the affected recall product. - Inform customers about this notice. - Identify and locate the Mega Ballast products concerned by this recall procedure. - Collect and put in quarantine the Mega Ballast products concerned by this recall procedure and then return them to your local distributor as per their return procedure. For questions or further assistance: Contact: Quality Department Email: QA@baltgroup.com Balt USA, LLC 29 Parker Suite 100 | Irvine CA, 92618 | USA Telephone: +949.788.1443 | Fax: +949.788.1444
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0188-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026