The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only performed under a physician's order. Recall
The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only performed under a physician's order.
FDA recall Class II
Ongoing
FDA lists this Class II recall for The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only performed under a physician's order.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0190-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 18, 2026
- FDA report date
- Oct 21, 2026
- Quantity in commerce
- 5395
- Distribution
- See key details
- FDA cause
- No Marketing Application
What is being recalled?
The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only performed under a physician's order.
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
UDI-DI: 05700693930923, 05700693930909 Software versions 3.6 and lower. ABL90 FLEX PLUS Instructions for Use versions prior to: 996-223, Version: 202606O. ABL90 FLEX Instructions for Use versions prior to: 996-189, Version: 202606L
Why was this product recalled?
In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges for some parameters, 3) Analyzer configuration, and 4) limitations and warnings.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Radiometer Medical ApS take?
On 8/18/2026, correction notices were mailed or emailed to customers who were asked to take the actions listed in the correction notice. The firm updated intended use and instructions for use. In addition, the firm is developing an updated software version that will modify the analyzer configuration in accordance with the updated FDA clearances. Customers are asked to do the following: 1) Implement the requested changes to the setup of your analyzer(s) as stated in the correction letter. 2) Review the changes to the labeling for the analyzers as stated in the updated Instructions for Use. 3) Complete and return the Recall Response Form via email to recall@radiometeramerica.com 4) If you are not the end user of the affected product, please ensure that this letter is distributed to the end user. If you have any questions, contact the firm at 1-800-736-0600 option 1, Monday - Friday 9am to 7pm (EST)
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0190-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology