Think Surgical Square Driver REF: 109126 Recall
Think Surgical Square Driver REF: 109126
FDA recall Class II
Ongoing
FDA lists this Class II recall for Think Surgical Square Driver REF: 109126. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0192-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 20, 2026
- FDA report date
- Oct 28, 2026
- Quantity in commerce
- 15 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Think Surgical Square Driver REF: 109126
Affected model & UDI checker
Search the exact identifiers published for Z-0192-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 109126 | Think Surgical Square Driver REF: 109126 | Check source notice | Not specified |
FDA code information
Lot: 25356/Primary DI: M9461091260
Why was this product recalled?
Robotic system square drive failed to meet design specifications potentially resulting in pin becoming stuck in square driver or the pin wobbling during drilling.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did THINK Surgical, Inc. take?
On August 25, 2026, THINK Surgical Inc. issued a Urgent Medical Device Recall Notification to affected consignees via email. Think Surgical sent an updated letter via email on September 22, 2026 to provide additional product information. THINK Surgical ask consignees to take the following actions: 1. Examine your inventory of square drivers for the lot number 25356. Quarantined return any inventory of that lot to THINK Surgical, Inc., 2. If you have distributed or transferred affected devices to other facilities or end users, please promptly notify them and provide a copy of this notification., 3. Fill out the accompanying Acknowledgement Form. Email the completed Acknowledgement Form to THINK Surgical, Inc at productreturns@thinksurgical.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0192-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.