Philips Smart-hopping 2.0 AP 1.4 GHz; Product/ Recall
Philips Smart-hopping 2.0 AP 1.4 GHz; Product/Model Number: 867216;
FDA recall Class II
Ongoing
FDA lists this Class II recall for Philips Smart-hopping 2.0 AP 1.4 GHz; Product/. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0199-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 16, 2026
- FDA report date
- Oct 28, 2026
- Quantity in commerce
- 174
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Philips Smart-hopping 2.0 AP 1.4 GHz; Product/Model Number: 867216;
Affected model & UDI checker
Search the exact identifiers published for Z-0199-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 867216 | Philips Smart-hopping 2.0 AP 1.4 GHz; Product/ | Check source notice | Not specified |
FDA code information
Product/Model Number: 867216; UDI-DI: N/A; System Serial Numbers: DK61624330, DK61623830, DK61724824, DK61623802, DK61623826, DK61624101, DK61624082, DK61623910, DK61624187, DK51313245, DK61724472, DK61724620, DK61724590, DK61724461, DK61724725, DK61724679, DK61724684, DK51313645, DK51313455, DK61724814, DK61724747, DK61624235, DK61422548, DK52316132, DK61724553, DK61724957, DK61724756, DK61523178, DK61724502, DK61724860, DK61523117, DK61724577, DK61724376, DK61724566, DK61422553, DK61322150, DK61523070, DK61523270, DK52617072, DK61523151, DK61422454, DK53817745, DK52617073, DK61422674, DK61422973, DK61523211, DK61322124, DK61523272, DK61523248, DK51313320, DK61422463, DK61523260, DK61322137, DK61322337, DK61322253, DK61322375, DK61322380, DK61422762, DK61422868, DK61322122, DK61422851, DK61322283, DK61523658, DK61422848, DK61422884, DK61422897, DK61523409, DK61422808, DK61422736, DK61422806, DK61422763, DK61523517, DK61422957, DK61523383, DK61523442, DK61322227, DK61422843, DK61523495, DK61423030, DK61422949, DK61322271, DK61322127, DK60621383, DK60621482, DK61322213, DK60321063, DK60320788, DK60621426, DK60621646, DK61322238, DK60621223, DK60621517, DK61322379, DK52316002, DK60320966, DK60621224, DK60621197, DK60821808, DK60621565, DK60821916, DK60821936, DK60320570, DK60320720, DK60320768, DK60320506, DK60320537, DK60321057, DK60320711, DK60320892, DK60320761, DK60320544, DK60321033, DK60320571, DK60320759, DK60320484, DK60321044, DK60320839, DK60320419, DK60320437, DK60320474, DK54820263, DK60320601, DK60320592, DK54820262, DK54820037, DK54819845, DK54820134, DK54819871, DK54820272, DK60320874, DK54519084, DK54519572, DK54820343, DK54819876, DK52516932, DK54519153, DK53817766, DK54518977, DK52516923, DK54519018, DK53817821, DK54820004, DK54819819, DK54820101, DK54819922, DK54518948, DK52516510, DK54519031, DK54518951, DK54519115, DK52516754, DK52516873, DK52516766, DK53817862, DK52516569, DK52516783, DK52516933, DK54418622, DK52516533, DK52315958, DK53217276, DK52316143, DK53217187, DK53217154, DK52316114, DK53217197, DK54418453, DK53918024, DK54418698, DK54418600, DK54418588, DK54418881, DK54519407, DK54418598;
Why was this product recalled?
Unreliable wireless roaming behavior of Smart-hopping 1.0 wireless clients during transport when connecting to Smart-hopping 2.0 infrastructure, resulting in loss of connectivity to the central monitoring system (PIC iX).
What actions did Philips North America take?
On september 16, 2026, URGENT Medical Device Correction, letters were sent to customers. Actions to be taken: 1. Direct clinical attendance of the patient during transport, when this issue is most likely to manifest, should be considered if not currently a part of the clinical workflow. In the event a persistent No Central Monit technical alarm is noted at the wireless device or a No Data technical alarm is noted at the central monitor; a restart of the wireless monitor may restore central monitoring when the patient is in known good wireless coverage area. 2. Pass this notice to all those who need to be aware within your organization or to any organization where affected product(s) have been potentially transferred. Actions planned by Philips to correct the problem 3. Philips representative will contact you to arrange the software upgrade SH 2.0 AP 1.4GHz which will correct this issue. If you need any further information, please contact your local Philips representative, or call (800) 722-9377.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0199-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology