Impella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump. Recall

Impella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Impella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0209-2027
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Sep 18, 2026
FDA report date
Nov 4, 2026
Quantity in commerce
179 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Impella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.

Affected model & UDI checker

Search the exact identifiers published for Z-0209-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedImpella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.Check source notice00813502010985
FDA code information

Product Code: 0042-0010-US, 0042-0010-UK, 0042-0010-EU, 0042-0010-IN; UDI: 00813502010985, 00813502010978, 00813502010961, 00813502011425; Serial No. IC6962, IC7004, IC7036, IC7170, IC7213, IC7214, IC7239, IC7241, IC7677, IC7787, IC7880, IC8105, IC8106, IC8295, IC8308, IC8315, IC8316, IC8335, IC8541, IC8543, IC11456, IC7937, IC9832, IC9833, IC5409, IC7798, IC7800, IC7946, IC8070, IC8337, IC8441, IC7825, IC8126, IC8630, IC8640, IC9804, IC11451, IC11497, IC8388, IC8751, IC9501, IC9989, IC10102, IC10103, IC10318, IC10589, IC10463, IC10979, IC11299, IC8164, IC9348, IC9500, IC9581, IC9648, IC9777, IC9793, IC9797, IC9915, IC9919, IC9922, IC10278, IC10302, IC10983, IC10984, IC10985, IC11174, IC11410, IC13231, IC13263, IC13264, IC13265, IC13681, IC13682, IC10040, IC10043, IC8401, IC8748, IC8980, IC8609, IC9178, IC9177, IC9324, IC9328, IC9557, IC9558, IC9744, IC9745, IC9746, IC9908, IC9921, IC9969, IC9970, IC10063, IC10064, IC10149, IC10150, IC10238, IC10239, IC10341, IC10366, IC10367, IC12407, IC12408, IC12409, IC13057, IC13060, IC13061, IC13064, IC8333, IC8494, IC12326, IC4952, IC5041, IC5158, IC5281, IC5283, IC5284, IC5403, IC5404, IC5413, IC5527, IC5588, IC5635, IC5636, IC5641, IC5643, IC5666, IC5667, IC5804, IC5805, IC5918, IC5921, IC6013, IC6343, IC6421, IC6422, IC6461, IC6463, IC6465, IC6777, IC6785, IC6944, IC6945, IC6946, IC7161, IC7233, IC7232, IC7235, IC7626, IC7738, IC7739, IC7792, IC7793, IC7826, IC7956, IC8012, IC8258, IC8423, IC8424, IC8499, IC8511, IC8723, IC8596, IC8665, IC8765, IC7827, IC7828, IC8404, IC11746, IC12344, IC12327, IC8426, IC13232, IC13698, IC7861, IC7862, IC8121, IC8222, IC8225.

View the original FDA recall notice

Why was this product recalled?

Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Abiomed, Inc. take?

Abiomed notified consignees on about 09/18/2026 via "URGENT MEDICAL DEVICE RECALL (REMOVAL)" letter. Consignees were informed of the issue and that affected units can continue to be used while awaiting service from Abiomed to ensure continuity of care. Consignees were instructed to identify if affected units are on hand, Abiomed will contact each consignee to coordinate the removal, update and return of affected devices, complete and return the provided business reply form by email to: Abiomed5510@sedgwick.com (or Fax to (866) 201-9339). If you have questions regarding returning this form, please contact (866) 482-2893. Inform all applicable personnel within your facility of the recall, and if any of the affected devices have been forwarded to another facility, to notify those facilities. If you have questions or concerns regarding this notice, please contact ra-abm-fieldaction@its.jnj.com or your local clinical field staff or call the Manager Product Issue Escalation & Field Actions at 978-646-1400.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology