Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: SWAN GANZ, KIT, CATALOG # PC37SGPHA Recall

Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: SWAN GANZ, KIT, CATALOG # PC37SGPHA

FDA recall Class I

Ongoing

FDA lists this Class I recall for Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: SWAN GANZ, KIT, CATALOG # PC37SGPHA. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0219-2027
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Sep 24, 2026
FDA report date
Nov 11, 2026
Quantity in commerce
112 kits
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: SWAN GANZ, KIT, CATALOG # PC37SGPHA

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

GTIN 382602846379 (ea) 50382602846374 (cs), LOT # 761270

View the original FDA recall notice

Why was this product recalled?

Kits contain infusion sets which were recalled for performance issues.

What actions did Cardinal Health 200, LLC take?

Cardinal Health issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 9/24/2026 via US Mail. The notice explained the issue and requested the discontinuation of the use of the affected BD Alaris Pump Infusion Sets as per the enclosed notice from Becton Dickinson. The notice instructed the following: If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: gmb-FieldCorrectiveAction@cardinalhealth.com For component credit or replacement, please contact the following: Hospitals Cardinal Customer Service at 866.551.0533 Distributors Distributor Management at 800.635.6021 Federal Government facilities Government Customer Service at 800.444.1166 Distributors were directed to notify their customers of the recall. For questions contact: Cardinal Health Quality Systems at 800.292.9332.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.