FDA RECALL RECORD · Z-0281-2026

RADPRO MOBILE, Mobile X-ray system Recall

RADPRO MOBILE, Mobile X-ray system

FDA recall Class II

Ongoing

FDA lists this Class II recall for RADPRO MOBILE, Mobile X-ray system. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0281-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 10, 2025
FDA report date
Oct 29, 2025
Quantity in commerce
24 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

RADPRO MOBILE, Mobile X-ray system

Affected model & UDI checker

Search the exact identifiers published for Z-0281-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedRADPRO MOBILE, Mobile X-ray systemCheck source notice08436046002166
FDA code information

UDI/DI 08436046002166, Serial Numbers: G30567, G31256, G32581, G30250, G28320, G30902, G29864, G35389, G30960, G33238, G31252, G33708, G30771, G33634, 31412, 31480, G33004, G33112, G33585, G31271, G31285, G32514, G32498, G35665.

View the original FDA recall notice

Why was this product recalled?

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

What actions did SEDECAL SA take?

Sedecal emailed their distributors on 06/10/2025 to remind them the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning. on 09/19/2025 the firm issued an Important Safety Notice to its direct consignees via email. This notice initiates Phase 2 of the recall strategy expanding the scope and providing information on stickers to cover gaps around the fine positioning control buttons and a DMC software update that stops unit movement if triggered by a short cicuit or prolonged button press with instructions. Consignees will be contacted by a Sedecal Customer support Service to coordinate the shipment of the necessary conversion kits (quantity and delivery address) to allow upgrading the affected units.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology