Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0286-2027

Orthoscan VERSA Accessory Cart, REF: 1000-0021 and Versa Mini C-Arm System, REF: 1000-0020 Recall

Orthoscan VERSA Accessory Cart, REF: 1000-0021 and Versa Mini C-Arm System, REF: 1000-0020

FDA recall Class II

Ongoing

FDA lists this Class II recall for Orthoscan VERSA Accessory Cart, REF: 1000-0021 and Versa Mini C-Arm System, REF: 1000-0020. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0286-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 3, 2026
FDA report date
Nov 4, 2026
Quantity in commerce
55 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Orthoscan VERSA Accessory Cart, REF: 1000-0021 and Versa Mini C-Arm System, REF: 1000-0020

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

UDI/ Mini C-Arm serial number (6Axxxx)/Accessory cart serial Number: (VCXXX): 1 6A0110 /VC0110 2 6A0111 /VC0111 3 6A0112 /VC0112 4 6A0115 /VC0113 5 6A0114 /VC0114 6 6A0117 /VC0115 7 6A0116 /VC0116 8 6A0113 /VC0117 9 6A0118 /VC0118 10 6A0119/ VC0119, VC0120; 11 6A0120 /VC0121; 12 6A0121 /VC0122; 13 6A0123 /VC0123; 14 6A0122 /VC0124; 15 6A0126 /VC0125; 16 6A0133 /VC0134; 17 6A0124 /VC0126; 18 6A0128 /VC0128; 19 6A0127 /VC0129; 20 6A0129 /VC0130; 21 6A0125 /VC0127; 22 6A0131 /VC0132; 23 6A0134 /VC0133; 24 6A0142 /VC0144; 25 6A0136 /VC0136; 26 6A0132 /VC0137; 27 6A0137 /VC0138; 28 6A0138 /VC0139; 29 6A0139 /VC0140; 30 6A0140 /VC0141; 31 6A0141 /VC0142; 32 6A0146 /VC0147; 33 6A0143 /VC0143; 34 6A0144 /VC0145; 35 6A0152 /VC0153; 36 6A0147 /VC0148; 37 6A0148 /VC0149; 38 6A0149 /VC0150; 39 6A0150 /VC0151; 40 6A0151 / VC0152; 41 6A0163 / VC0165; 42 6A0130 / VC0131; 43 6A0155 / VC0156; 44 6A0157 / VC0157; 45 6A0156/ VC0158; 46 6A0158 / VC0159, VC0160; 47 6A0159/ VC0161; 48 6A0160/ VC0162; 49 6A0161 / VC0163; 50 6A0162 VC0164; 51 6A0153 VC0154; 52 6A0164 VC0166; 53 6A0165 VC0167

View the original FDA recall notice

Why was this product recalled?

The accessory cart weldment does not conform to weld dimensional specifications which may result in separation of the monitor arm mounting structure from the cart.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did ZIEHM-ORTHOSCAN INC take?

On September 11, 2026, Ziehm-Orthoscan issued a Urgent: Medical Device Recall Notification via Email. Ziehm-Orthoscan ask consignees to take the following actions: 1. The device may continue to be used until the corrective action has been completed, provided the user follows the precautions described in this notification and no signs of instability or unintended movement are observed. 2. Users should exercise caution when positioning, moving, or operating the system and monitor assembly, avoid fully extending the display arm until the corrective action has been performed. 3. If unintended movement, instability, loosening, or any other condition related to the issue described in this notification is observed, discontinue use of the affected system and contact Ziehm- Orthoscan immediately for further instructions. 4. Please ensure that a copy of this notification is made available to all users and operators of the affected device. It is recommended that this notification be posted in a visible location near the device until the corrective action has been completed. 5. If the affected device has been transferred, sold, or relocated to another facility or department, please ensure that a copy of this notification is provided to the current user or owner of the device. 6. Please maintain awareness of this notification until the corrective action has been completed and verified by Ziehm-Orthoscan or its authorized representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology