Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0287-2027

Generic Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery System Recall

Generic Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery System Model/Catalog Number: 9880CM29A Software Version: Not applicable. Product Description: Prosthesis, mitral valve, percutaneously delivered Component: No

FDA recall Class I

Ongoing

FDA lists this Class I recall for Generic Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery System. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0287-2027
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Oct 5, 2026
FDA report date
Nov 11, 2026
Quantity in commerce
164
Distribution
See key details
FDA cause
Process change control

What is being recalled?

Generic Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery System Model/Catalog Number: 9880CM29A Software Version: Not applicable. Product Description: Prosthesis, mitral valve, percutaneously delivered Component: No

Affected model & UDI checker

Search the exact identifiers published for Z-0287-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
9880CM29AGeneric Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery SystemCheck source noticeNot specified
FDA code information

Lot Code: Model No.: 9880CM29A; UDI-DI: 00690103219777; Recalled Lot Number(s): 67718561, 67804014, 67887043, 67887044.

View the original FDA recall notice

Why was this product recalled?

Due to a manufacturing issue which may result in cardiovascular balloon rupturing.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Edwards Lifesciences, LLC take?

On 10/05/2026, an "URGENT: MEDICAL DEVICE RECALL (PRODUCT REMOVAL)" Letter was sent via various delivery methods (e.g., electronic delivery (e-mail), FedEx Priority Overnight mail, or on-site visit) to customers informing them that, due to complaints certain manufacturing lots of the SAPIEN M3 Commander M delivery systems potentially contain a manufacturing defect, which may result in balloon burst during use. Customers are instructed to: " Review this notice carefully and share it with the appropriate clinical staff at their site. " Immediately examine their inventory and quarantine units subject to this recall. Do not use these affected units. Units with other lot numbers are not impacted by this recall and can continue to be used. " Complete the attached Customer Acknowledgment Form and return it to Edwards TMTT Representative or via email to US_FCA@edwards.com " Contact Customer Service at 1-800-424-3278 or email: CustomerServ_TMTT@edwards.com to obtain a Returned Goods Authorization (RGA) number and return any impacted product to Edwards. Edwards asks that customers complete the last two actions listed above within 15 days of receipt of this notification. Customers' assistance is crucial to ensuring this recall is executed thoroughly and promptly. This notice has been communicated by Edwards to the appropriate Regulatory Authorities. For questions and further assistance, contact an Edwards representative who can assist with product return.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology