FDA RECALL RECORD · Z-0301-2027

EMAG system Recall

EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated

FDA recall Class II

Ongoing

FDA lists this Class II recall for EMAG system. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0301-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 11, 2026
FDA report date
Nov 4, 2026
Quantity in commerce
33
Distribution
See key details
FDA cause
Software change control

What is being recalled?

EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated

Affected model & UDI checker

Search the exact identifiers published for Z-0301-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
418591/418591UEMAG systemCheck source noticeNot specified
FDA code information

UDI-DI 357302BUDI000544TP Serial Numbers IM03343 IM03283 IM03104 IM03329 IM03376 IM03474 IM03321 IM03071 IM03175 IM03194 IM03515 IM03516 IM03240 IM03478 IM03479 IM03482 IM03034 IM03092 IM03378 IM03414 IM03177 IM03193 IM03506 IM3054 IM03439 IM03350 IM03380 IM03131 IM03148 IM03153 IM03275 IM03314 IM03351

View the original FDA recall notice

Why was this product recalled?

Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2

What actions did BioMerieux SA take?

The firm began notifying consignees on September 11, 2026. Customers were informed they may continue use of EMAG systems and there is no risk of false results. In case of 10003 error heater time-out during an extraction protocol, customers are instructed to reset the system to run the protocol again to prevent delay/loss of sample. The company will contact customers to schedule a bioMerieux Field Service Engineer (FSE) visit to update affected EMAG systems to the previous firmware patch SP4 Patch 1 to fix the issue.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology