Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0306-2027

PENTAX Medical Ultrasound Endoscope. Model Number. EG36-J10UR. Recall

PENTAX Medical Ultrasound Endoscope. Model Number. EG36-J10UR.

FDA recall Class II

Ongoing

FDA lists this Class II recall for PENTAX Medical Ultrasound Endoscope. Model Number. EG36-J10UR.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0306-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 9, 2026
FDA report date
Nov 4, 2026
Quantity in commerce
84 units
Distribution
See key details
FDA cause
Unknown/Undetermined by firm

What is being recalled?

PENTAX Medical Ultrasound Endoscope. Model Number. EG36-J10UR.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Model No. EG36-J10UR; UDI-DI: 04961333239573, 04961333239580, 04961333239597, 04961333239603, 04961333239627, 04961333248919, 04961333248926, 04961333248933, 04961333248957, 04961333252756; All Serial Numbers.

View the original FDA recall notice

Why was this product recalled?

Warnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and during the procedure.

What actions did Pentax of America Inc take?

An Urgent Field Safety Notice dated 09/02/2026 was mailed to consignees. The notice informs consignees of two updates to the IFU of PENTAX Medical products, specifically the ultrasound endoscope devices and water bottle assembly. Consignees are to inform their staff of this notice and include the warning and updates to their IFUs. Additionally, consignees are instructed to change water for each procedure and fill the bottle approximately 2/3 full, or to the 200 marking on the bottle scale, with sterile water. If the water/balloon feeding function does not work consignees are to fill the water bottle with deaerated sterile water. PENTAX Medical can be contacted by phone at 1-800-431-5880 Monday through Friday from 8:30 AM EST to 5:00 PM EST, or by email at customeradvisories@pentaxmedical.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.