Magseed Pro 12 cm soft tissue marker. Recall
Magseed Pro 12 cm soft tissue marker. Model Number: MS2-17-1-12. Product Description: The Endomag Magseed Pro Magnetic Marker System is a sterile, single use device composed of a marker preloaded in a 17ga Needle Delivery System. The Endomag Magseed Pro Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Magseed Pro 12 cm soft tissue marker.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0317-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 1, 2025
- FDA report date
- Nov 5, 2025
- Quantity in commerce
- 540 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Magseed Pro 12 cm soft tissue marker. Model Number: MS2-17-1-12. Product Description: The Endomag Magseed Pro Magnetic Marker System is a sterile, single use device composed of a marker preloaded in a 17ga Needle Delivery System. The Endomag Magseed Pro Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures.
Affected model & UDI checker
Search the exact identifiers published for Z-0317-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MS2-17-1-12 | Magseed Pro 12 cm soft tissue marker. | Check source notice | Not specified |
| Not specified | Magseed Pro 12 cm soft tissue marker. | Check source notice | 15060391210466 |
FDA code information
Model No. MS2-17-1-12. GTIN:15060391210466. UDI: 15060391210466250327-01, 15060391210466250415-07, 15060391210466250523-10, 15060391210466250612-50. Batch Number: 250327-01, 250415-07, 250523-10, 250612-50.
Why was this product recalled?
Potential for contamination with cotton fibers.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Endomagnetics Ltd., take?
Endomagnetics, Ltd notified consignees on about 10/01/2025 via certified mail and email through 3rd party IQVIA. Consignees were instructed to examine their inventory and cease use of those units, quarantine any affected units, complete and return the response form, forward the notification to all affected personnel and customers if further distributed, and post a copy of the notice in a visible area for awareness and keep a copy for your records.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0317-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology