Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0324-2026

SpineAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SPR.1.2.0, orthopedic augmented reality Recall

SpineAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SPR.1.2.0, orthopedic augmented reality

FDA recall Class II

Terminated

FDA lists this Class II recall for SpineAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SPR.1.2.0, orthopedic augmented reality. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0324-2026
Recall classification
Class II
Enforcement status
Terminated
Initiation date
Sep 16, 2025
FDA report date
Nov 5, 2025
Quantity in commerce
3 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

SpineAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SPR.1.2.0, orthopedic augmented reality

Affected model & UDI checker

Search the exact identifiers published for Z-0324-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedSpineAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SPR.1.2.0, orthopedic augmented realityCheck source notice00860003884126
FDA code information

UDI/DI 00860003884126

View the original FDA recall notice

Why was this product recalled?

A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

What actions did Surgical Theater Inc take?

Surgical Theater issued an URGENT: Medical Device Software correction notice dated September 17, 2025 to its consignees on 09/22/2025 via email. The notice explained the issue, risk to health, and provided the following: Customers can continue to use the software version(s) noted above until a software patch is made available by Surgical Theater. Until the software patch is installed, please follow these steps: 1. Read and share this notice with all relevant staff. 2. Continue to use the 3D Fusion functions to review the fusion results for all case files to verify alignment of secondary layer(s) to primary layer and to adjust fusion as necessary, and 3. Acknowledge your receipt of this notification by completing and returning the included acknowledgement form via email to quality@surgicaltheater.com. A software patch is currently undergoing verification and should be available in the next 10 days. Surgical Theater has revised the 2D Image Fusion function to ensure the x, y and z axis rotation values are consistently applied when generating the aligned Axial, Sagittal and Coronal images. In addition, the rotate left/right buttons in all three planes were modified so the volume rotation stays consistent with the respective plane orientation. A Surgical Theater representative will contact your facility to schedule installation once the patch is available. For questions regarding this notification contact: Kevin Murrock, Sr. Director of Quality & Regulatory, Tel. 330-472-6520 (Monday through Friday, 9:00 AM to 5:00 PM, Eastern Time), email: kmurrock@surgicaltheater.com

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

  5. Recall termination recorded

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology