Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0331-2027

Halyard MINOR PACK. Recall

Halyard MINOR PACK. Model/Catalog Number: VASF105-08.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Halyard MINOR PACK.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0331-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 8, 2026
FDA report date
Nov 11, 2026
Quantity in commerce
108 kits
Distribution
See key details
FDA cause
Process control

What is being recalled?

Halyard MINOR PACK. Model/Catalog Number: VASF105-08.

Affected model & UDI checker

Search the exact identifiers published for Z-0331-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
VASF105-08Halyard MINOR PACK.Check source noticeNot specified
FDA code information

Model/Catalog Number: VASF105-08. UDI-DI: 10809160451960. Lot Number: 1659384. Expiration Date: 9/30/2027.

View the original FDA recall notice

Why was this product recalled?

Potential for open seals on header bags, compromising sterility.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did AVID Medical, Inc. take?

AVID Medical notified consignees on about 09/08/2026 via emailed "URGENT: MEDICAL DEVICE CORRECTION" letter. Consignees were instructed to Segregate, Quarantine, and Inspect all kits identified. If the header bag seal appears to be compromised, discard the entire kit. Consignees were also requested to complete and return the provided Recall Response Form. Distributors were instructed to notify customers if affected kits were further distributed. On 10/06/2026 AVID Medical sent an updated "URGENT: MEDICAL DEVICE CORRECTION" letter to consignees via email. The strategy was updated and consignees were instructed not to use the affected kits and that they should either discard or return them. If returning, they should contact customer service for a Return Goods Authorization. Consignees were again requested to complete and return the provided Recall Response Form. Distributors were again instructed to notify customers if affected kits were further distributed.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology