Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0347-2027

Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60) Recall

Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),

FDA recall Class II

Ongoing

FDA lists this Class II recall for Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0347-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 30, 2025
FDA report date
Nov 11, 2026
Quantity in commerce
118 unts.
Distribution
See key details
FDA cause
Process control

What is being recalled?

Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Serial number/Shimadzu Medical Systems #/UDI: M1F0D3BF4001/ SMS20250167/ (01)04540217067145(11)250; M1F0CC4EB001/ SMS20250098/ (01)04540217067022(11)241; M1F0D27EA001/ SMS20250001/ (01)04540217067138(11)241; 41F0F43E7001/ SMS20250034/ (01)04540217068951(11)240; M1F0D3BE7001/ SMS20250248/ (01)04540217067145(11)240; M1F0D3BE6001/ SMS20240196/ (01)04540217067145(11)240; 41F0F40E5001/ N/A/ (01)04540217068937(11)240; 41F0F41E5001/ N/A/ (01)04540217068944(11)240; 41F0F43E5001/ SMS20240236/ (01)04540217068951(11)240; M1F0CD7E4001/ SMS20240162/ (01)04540217067046(11)240; 41F0F41E1001/ SMS20240062/ (01)04540217068944(11)240; 41F0F43DB002/ SMS20240101/ (01)04540217068951(11)231; 41F0F43DB001/ SMS20240100/ (01)04540217068951(11)231; M1F0CD7D4001/ SMS20230178/ (01)04540217067046(11)230

View the original FDA recall notice

Why was this product recalled?

The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION take?

On September 30, 2026 Shimadzu Corp. issued a Urgent Medical Device Recall Notification to affected consignees via email. Shimadzu asked consignees to take the following actions; 1. If the bolts become loose, an abnormal noise will occur when moving the C-arm. Please check for abnormal noises during pre-study inspections. 2. If an abnormal noise is heard, please stop using the product and contact our service department. 3. Please complete the Medical Device Recall Customer Acknowledgement.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology