Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0349-2027

Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Recall

Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, single use devices intended for long term implantation under the skin in the mastoid region on either side of the head. The cochlear implant system works by converting sound in the environment into electrical code. It transmits this code to the auditory nerve and on to the brain where it is interpreted as sound. The Cochlear Implant System consists of internal (implanted) and external components. The implanted component consists of a receiver/stimulator and an electrode array. The external components consist of a sound processor and their associated transmitting coils and cables and must be worn for the implant to work.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0349-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 30, 2026
FDA report date
Nov 11, 2026
Quantity in commerce
4 device in the U.S.
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, single use devices intended for long term implantation under the skin in the mastoid region on either side of the head. The cochlear implant system works by converting sound in the environment into electrical code. It transmits this code to the auditory nerve and on to the brain where it is interpreted as sound. The Cochlear Implant System consists of internal (implanted) and external components. The implanted component consists of a receiver/stimulator and an electrode array. The external components consist of a sound processor and their associated transmitting coils and cables and must be worn for the implant to work.

Affected model & UDI checker

Search the exact identifiers published for Z-0349-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
CI1000Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576Check source noticeNot specified
FDA code information

Model: CI1000 Product/Catalog: P1979576 UDI-DI code: 9321502069847 Serial Numbers: 1020210051972 1020210070461 1020210145046 1020210048286

View the original FDA recall notice

Why was this product recalled?

Due to a manufacturing deviation of incomplete electrode welds of cochlear implants and resulting in one or more open circuits

What actions did Cochlear Limited take?

On 09/30/2026, an "URGENT Field Safety Corrective Action Notification" was sent to U.S. customers informing them that a potential condition has been identified that could result in one or more open circuits affecting individual electrode pathways with the implant. Customers were instucted to not implant and return affected products to Cochlear Limited. For questions contact CNA_Quality@cochlear.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology