Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040. Recall

Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0356-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 2, 2026
FDA report date
Nov 11, 2026
Quantity in commerce
22 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Part Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921.

View the original FDA recall notice

Why was this product recalled?

Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did CORIN MEDICAL, LTD. take?

Corin notified consignees on about 10/02/2026 via emailed "URGENT MEDICAL DEVICE REMOVAL" letter. Consignees were instructed to identify the products present in inventory and quarantine them, return the devices to Corin (replacement parts will be provided), provide the notification letter to all involved parties, and complete and return the acknowledgement form. Additionally, Corin requested notification if affected units have been further distributed as well as any affected customers so that they may be notified.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology