FDA RECALL RECORD · Z-0359-2027

LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303 Recall

LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303

FDA recall Class II

Ongoing

FDA lists this Class II recall for LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0359-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 5, 2026
FDA report date
Nov 11, 2026
Quantity in commerce
30,168 test cartridges (2,514 Kits)
Distribution
See key details
FDA cause
Error in labeling

What is being recalled?

LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

All lots UDI-DI pending

View the original FDA recall notice

Why was this product recalled?

Instructions for Use updated to include explicit guidance that repeat testing is recommended whenever a positive family-level only result is obtained. Issue may contribute to incorrect or delayed diagnosis of bloodstream infection.

What actions did Luminex Corporation take?

On October 5, 2026, the firm began notifying customers via "Urgent Field Safety Notice - Medical Device Recall" letters. Consignees were informed that the current Instructions for Use for the Gram-Negative Blood Culture Assay do not adequately describe interpretation of certain positive family-level results. Specifically, when a positive family-level only result is obtained ("family-level only result" refers to a positive organism family-level detection without an accompanying genus- or species-level identification), repeat testing with the LIAISON PLEX Gram-Negative Blood Culture Assay is recommended. Absence of this instruction may increase the likelihood that certain reportable detections are not appropriately recognized by the user and may contribute to an incorrect interpretation of test results. Customers were instructed to implement the following interpretation guidance: When a positive family-level only result is obtained, repeat testing with the LIAISON PLEX BCN assay is recommended. DiaSorin is updating the Instructions for Use to include the revised guidance. Revised labeling will be provided following implementation of the labeling revision. Please contact Technical Support Services with any questions or concerns at the following phone numbers:1-877-785-2323 (U.S. and Canada); +1-512-381-4397 (Outside U.S. and Canada); 1-512-219-5114 (Fax) and email:support@luminexcorp.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.