Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0371-2026

All Serial Numbers/BD Pyxis CII Safe ES Tower Main, REF: 1116-00 Recall

All Serial Numbers/BD Pyxis CII Safe ES Tower Main, REF: 1116-00

FDA recall Class II

Ongoing

FDA lists this Class II recall for All Serial Numbers/BD Pyxis CII Safe ES Tower Main, REF: 1116-00. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0371-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 30, 2025
FDA report date
Nov 5, 2025
Quantity in commerce
443 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

All Serial Numbers/BD Pyxis CII Safe ES Tower Main, REF: 1116-00

Affected model & UDI checker

Search the exact identifiers published for Z-0371-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1116-00All Serial Numbers/BD Pyxis CII Safe ES Tower Main, REF: 1116-00Check source noticeNot specified
FDA code information

Ref #/UDI: 1116-00 10885403512605

View the original FDA recall notice

Why was this product recalled?

Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did CareFusion 303, Inc. take?

On September 30, 2025, BD issue a Urgent Medical Device Recall Expansion notification to affected consignees via FedEx. BD is asking consignees to take the following actions: 1. Affected products can continue to be used per existing Manuals and Guides. 2. In the event of an accidental fluid spill on the device or in a drawer, users should following instructions outlined within the customer notification. 3. Review the customer letter attachment which provides warnings, cautions and recommendations regarding fluid ingress. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology