Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0374-2026

CODMAN Disposable Perforator 9mm. Cranial Perforator. Recall

CODMAN Disposable Perforator 9mm. Cranial Perforator.

FDA recall Class I

Ongoing

FDA lists this Class I recall for CODMAN Disposable Perforator 9mm. Cranial Perforator.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0374-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Sep 26, 2025
FDA report date
Nov 12, 2025
Quantity in commerce
5,109 units
Distribution
See key details
FDA cause
Process change control

What is being recalled?

CODMAN Disposable Perforator 9mm. Cranial Perforator.

Affected model & UDI checker

Search the exact identifiers published for Z-0374-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
261223CODMAN Disposable Perforator 9mm. Cranial Perforator.Check source noticeNot specified
Not specifiedCODMAN Disposable Perforator 9mm. Cranial Perforator.Check source notice10381780513605
FDA code information

Model No. 261223; UDI: 10381780513605; Lot No. 5828553, 6264478, 6264481, 6264482, 7230672, 7230673, 7230674, 7230675, 7261971, 7268008, 7268009, 7268010, 7268011, 7268012, 7268013, 7268014, 7268015, 7268016, 7268017, 7268018, 7268019, 7282954, 7282955, 7293422, 7293423, 7293424, 7293425, 7293426, 7293427, 7293428, 7300667, 7300668, 7300669, 7300670, 7300672, 7300673, 7300674, 7371082, 7371083, 7379652, 7379922, 7379923, 7379924, 7379926, 7379928, 7379660, 7379925, 7379927, 7450559, 7450560, 7450561, 7450562, 7450563, 7450564, 7450565, 7450566, 7451273, 7451274, 7451275, 7451276, 7451277, 7451278, 7451279; Expiration Date: 2024-03-31 to 2029-07-31.

View the original FDA recall notice

Why was this product recalled?

Expansion of recall of size 14 mm Codman Disposable Perforators for inadequate welds to size 11 mm and 9 mm Codman Disposable Perforators out of an abundance of caution.

What actions did Integra LifeSciences Corp. (NeuroSciences) take?

An URGENT: MEDICAL DEVICE RECALL notification dated 9/26/25 was mailed to consignees. Customers with affected devices are to be immediately removed and quarantined. Customers are to send their completed Acknowledgement Form to FCA3@integralife.com or by fax at 1-609-750-4220. Forward the recall notification to those who use the product for their awareness. Integra Customer Service will reach out to consignees to facilitate product return. Distributors and Sales Representatives with affected devices are to remove product from further distribution. They are to send their completed Acknowledgement Form to FCA3@integralife.com or by fax at 1-609-750-4220. Distributors are to create an acknowledgement form from the distributor to their customers and provide them with a copy of the recall notification. Sales Representatives are to receive further directions from Integra Customer Service after their Acknowledgement Form has been received. Consignees with any questions are to call 1-800-654-2873 or email custsvcnj@integralife.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology