Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control Recall
Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OCRD-202, OCRD-303, OCRD-SP, OCRD-MP, OCRD-UL; Product Description: In Vitro Diagnostic external control
FDA recall Class II
Ongoing
FDA lists this Class II recall for Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0415-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 1, 2025
- FDA report date
- Nov 5, 2025
- Quantity in commerce
- 4462
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OCRD-202, OCRD-303, OCRD-SP, OCRD-MP, OCRD-UL; Product Description: In Vitro Diagnostic external control
Affected model & UDI checker
Search the exact identifiers published for Z-0415-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| OCRD-101 | Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control | Check source notice | Not specified |
| OCRD-202 | Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control | Check source notice | Not specified |
| OCRD-303 | Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control | Check source notice | Not specified |
| Not specified | Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control | Check source notice | 00884883011946 |
| Not specified | Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control | Check source notice | 00884883011953 |
| Not specified | Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control | Check source notice | 00884883011960 |
| Not specified | Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control | Check source notice | 00884883011977 |
| Not specified | Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control | Check source notice | 00884883011984 |
| Not specified | Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control | Check source notice | 00884883011991 |
| Not specified | Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control | Check source notice | 00884883012004 |
FDA code information
OCRD-UL: UDI/DI 00884883011946, Lot Number OCRD2704U, Exp 2027-04-30; OCRD-L: UDI/DI 00884883011953, Lot Number OCRD2704L, Exp 2027-04-30; OCRD-101: UDI/DI 00884883011960, Lot Number OCRD27041, Exp 2027-04-30; OCRD-202: UDI/DI 00884883011977, Lot Number OCRD27042, Exp 2027-04-30; OCRD-303: UDI/DI 00884883011984, Lot Number OCRD27043, Exp 2027-04-30; OCRD-SP: UDI/DI 00884883011991, Lot Number OCRD2704S, Exp 2027-04-30; OCRD-MP: UDI/DI 00884883012004, Lot Number OCRD2704M, Exp 2027-04-30.
Why was this product recalled?
The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays.
What actions did Microgenics Corporation take?
ThermoFisher Scientific issued an URGENT -MEDICAL DEVICE RECALL notice to its consignees on 10/02/2025 via email. The notice explained the issue, risk to health, and requested the following: "ACTIONS TO BE TAKEN BY A USER 1. If you are a user of the High sensitivity Troponin I Assay, determine if you have inventory of the affected lots of MAS Omni"CARDIO" control as listed in Table 1. As stated in the instructions for use, instrument values provided are specific to this lot of control only and are intended to assist the laboratory in establishing its own means and ranges. 2. You may continue to use controls that meet your laboratory s performance requirements. 3. If the variability of the material exceeds your laboratory s specification, dispose of the material in accordance with local, regional, and national regulations. After disposing of the material, contact TechService.mgc@thermofisher.com for a credit. 4. Customers using the affected product with standard/ regular Troponin I assays, can continue to use the product as long as it meets their laboratory s specification and continue monitoring performance of Troponin I in MAS Omni"CARDIO" control. 5. This recall notice needs to be passed to anyone within your organization or any organization where the potentially affected lots have been transferred. 6. Retain a copy of this letter for your laboratory records. 7. Please complete the Recall Response Form attached to this Notification within 10 days of the date of this letter and return to Thermo Fisher as instructed in the form to E-mail:MGCFSCA@thermofisher.com 8. If you have any additional questions regarding this Recall Notice, contact Thermo Fisher Scientific Technical Support at toll free 1-800-232-3342 (USA and Canada only) or 510-979-5000 (International), Option 2, then option 3 or by email: TechService.mgc@thermofisher.com." "ACTIONS TO BE TAKEN BY A DISTRIBUTOR 1. Please notify your customers of this Recall Notice using this recall notice and request t
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0415-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology