Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0416-2026

Brand Name: Sentinel V11 Product Name: Sentinel V11 Recall

Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology Information Management System Component: No The Sentinel Cardiology Information Management System is intended to connect to supported medical devices and analyzers in order to download, store, access and manage cardiovascular information and to manage patient and facility information.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Sentinel V11 Product Name: Sentinel V11. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0416-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 16, 2025
FDA report date
Nov 12, 2025
Quantity in commerce
142
Distribution
See key details
FDA cause
Software design

What is being recalled?

Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology Information Management System Component: No The Sentinel Cardiology Information Management System is intended to connect to supported medical devices and analyzers in order to download, store, access and manage cardiovascular information and to manage patient and facility information.

Affected model & UDI checker

Search the exact identifiers published for Z-0416-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
92810Brand Name: Sentinel V11 Product Name: Sentinel V11Check source noticeNot specified
Not specifiedBrand Name: Sentinel V11 Product Name: Sentinel V11Check source notice10841522125751
FDA code information

Catalog Number: 92810 UDI-DI code; (01)10841522125751 Software version: 11.6.0

View the original FDA recall notice

Why was this product recalled?

Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record

What actions did Spacelabs Healthcare, Ltd. take?

On 09/15/2025, the firm sent via email an "URGENT MEDICAL DEVICE CORRECTION" Letter to customer informing them of two issues associated with the Sentinel Software Version 11.6.0. The issues are: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record. Customers are instructed to: 1. Print and Post Notification: Please print the related Urgent User Notification (Appendix A) in all location(s) where Sentinel 11.6.0 is used to configure ABP and/or Ambulatory ECG (Holter) devices. 2. Notify all relevant employees within their organization, including satellites, of this recall 3. Confirm the above action has been taken: Go to the QR code below (Appendix B) and confirm that they have read, understood, and posted the user notification at their workstations. 4. Contact Customer Support team at gtsdc@spacelabs.com with any questions regarding this identified issue. A fix for these issues is being developed, and is anticipated to be available by the end of September 2025. A Spacelabs Healthcare representative will notify customers once the fix is available and will work with customers to schedule the software update. Note, rolling back Sentinel 11.6.0 to a previous version is not a valid option.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.