TruDi Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01; Recall
TruDi¿ Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01;
FDA recall Class II
Ongoing
FDA lists this Class II recall for TruDi Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0420-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 3, 2025
- FDA report date
- Nov 12, 2025
- Quantity in commerce
- 1198 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
TruDi Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01;
Affected model & UDI checker
Search the exact identifiers published for Z-0420-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | TruDi Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01; | Check source notice | 10846835021103 |
FDA code information
Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01; UDI-DI: 10846835021103;
Why was this product recalled?
It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy.
What actions did Integra LifeSciences Corp. (NeuroSciences) take?
On October 3, 2025 URGENT: VOLUNTARY MEDICAL DEVICE RECALL letters were sent to customers. Actions to be Taken by Customers 1. Examine the your system to determine if you have product on hand subject to this field recall (Rev 00 and Rev01). 2. If you have revision 00 and/or revision 01, please check the box on the enclosed Acknowledgement Form I do have affected product(s) and record the total quantity of the affected product that you have. 3. If you do not have affected product, check the box, I do not have affected product. 4. Complete the rest of the Acknowledgement Form and return. 5. Keep a copy of the form for your records. 6. Forward this notice to those who utilize the product so they are aware of this recall and can identify and quarantine any affected product that may remain in clinical areas. 7. When your form is received, and it is noted that you have Rev 00 or Rev 01 products, Customer Service will contact you and provide a Return Material Authorization (RMA) number and directions to return the affected product. You can request a replacement which you will receive depending on availability. 8. In an effort to prioritize replacement Adapters please note below (in the Medical Facilities Acknowledgement Form) if you have Rev 02 and Rev 03 to use for your upcoming planned surgeries.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0420-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026