DigitalDiagnost 4 (1) High Performance, (2) Flex/Value; Recall
DigitalDiagnost 4 (1) High Performance, (2) Flex/Value; Model Number: (1) 712031, (2) 712032;
FDA recall Class II
Ongoing
FDA lists this Class II recall for DigitalDiagnost 4 (1) High Performance, (2) Flex/Value;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0464-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 1, 2025
- FDA report date
- Nov 19, 2025
- Quantity in commerce
- 12 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
DigitalDiagnost 4 (1) High Performance, (2) Flex/Value; Model Number: (1) 712031, (2) 712032;
Affected model & UDI checker
Search the exact identifiers published for Z-0464-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 712031 | DigitalDiagnost 4 (1) High Performance, (2) Flex/Value; | Check source notice | Not specified |
| 712032 | DigitalDiagnost 4 (1) High Performance, (2) Flex/Value; | Check source notice | Not specified |
| Not specified | DigitalDiagnost 4 (1) High Performance, (2) Flex/Value; | Check source notice | 00884838065321 |
| Not specified | DigitalDiagnost 4 (1) High Performance, (2) Flex/Value; | Check source notice | 00884838074514 |
FDA code information
1. Model Number: 712031; UDI-DI: 00884838065321; Serial Numbers: 10001181, 10001182, 10001185, 10001186, 10001188, 10001189, SN16000022, SN20000013; 2. Model Number: 712032; UDI-DI: 00884838074514; Serial Numbers: 10001179, 19000341, 19000342, 19000343;
Why was this product recalled?
Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Philips Medical Systems DMC GmbH take?
On October 1, 2025, "URGENT MEDICAL DEVICE CORRECTION" letters dated September 30, 2025, were sent to customers. The letters described the problem, product and actions to be taken. Actions to be taken: It is safe to continue to use the system(s) in accordance with the intended use. It is recommended that customers use the motorized movement option until the technical solution has been implemented. The system may experience motor drive errors if the force for movement exceeds 110N, which restricts the motorized movement. If the motorized movement does not function, customers should contact their local service representative. B. Circulate this notice to all users of this device so that they are aware of the issue and associated hazard/harm. C. Please retain this Urgent Medical Device Correction letter with your system(s) until the solution is installed; ensure the notice is in a place likely to be seen/viewed. 5. Actions planned by Philips to correct the problem. Philips will provide the technical solution details when finalized. This notice has been reported to the appropriate Regulatory Agencies. If you need any further information or support concerning this issue, please contact your local Philips representative. For North America, contact the Customer Care Solutions Center (1-800-722-9377, 8AM 8PM EST, Monday-Friday).
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0464-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026